FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 15073410
·
Received July 21, 2022
Report
- Report Number
- 3013756811-2022-75299
- Event Type
- Malfunction
- Date Received
- July 21, 2022
- Date of Event
- July 2, 2022
- Report Date
- July 2, 2022
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613793
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION.
Additional Manufacturer Narrative · 0
THE FAILURE INVESTIGATION HAS BEEN COMPLETED AND THE ALLEGED OCCLUSION ISSUE WAS VERIFIED IN THE PUMP LOGS; HOWEVER, NO FAILURE WAS IDENTIFIED. ADDITIONALLY, BASED ON THE ANALYSIS, THE ALLEGED FILL ESTIMATE ISSUE COULD NOT BE VERIFIED.
Description of Event or Problem · 0
IT WAS REPORTED THAT AN OCCLUSION ALARM OCCURRED. CUSTOMER CHANGED PUMP SUPPLIES TO ADDRESS THE ISSUE. ADDITIONALLY, IT WAS REPORTED THAT THE INSULIN GAUGE WAS INACCURATE. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS IN 150-153 MG/DL RANGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1698765 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1000354 | 00850006613793 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | Female | INSULIN TYPE: HUMALOG |