FDA Adverse Event Malfunction Summary report: N

OMNIPOD 5 POD

MDR report key: 15048336 · Received July 18, 2022

Report

Report Number
3004464228-2022-12173
Event Type
Malfunction
Date Received
July 18, 2022
Date of Event
July 13, 2022
Report Date
July 14, 2022
Manufacturer
INSULET CORPORATION
Product Code
QFG
PMA / PMN Number
K203768
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ACCORDING TO THE COMPLAINANT THE DEVICE WILL NOT BE RETURNED FOR INVESTIGATION. IT WAS REPORTED THAT THE CANNULA HAD DISLODGED FROM THE INFUSION SITE. THIS CONDITION COULD INTERRUPT INSULIN DELIVERY AND CONTRIBUTE TO HYPERGLYCEMIA. LOT RELEASE RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. SPECIFICALLY, A POD IS PAIRED TO A PDM AND PUT THROUGH SIMULATED USE TESTING INCLUDING COMMUNICATING WITH THE PDM, DEPLOYMENT, DELIVERING FLUID, OCCLUSION DETECTION, AND FREEDOM FROM HAZARD ALARMS.

Additional Manufacturer Narrative · 0

CORRECTION TO D1 - BRAND NAME CHANGED FROM UNAVAILABLE TO OMNIPOD 5 POD. D2A - COMMON DEVICE NAME CHANGED FROM UNAVAILABLE TO ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP. D2B - PROCODE CHANGED FROM UNAVAILABLE TO QFG. CORRECTION TO D(4): MODEL NO CHANGED FROM PT-000536 TO PT-000435. CATALOG NO CHANGED FROM UNAVAILABLE TO POD-BLE-H1-520. UNIQUE IDENTIFIER (UDI) # CHANGED TO (B)(4). CORRECTION TO G(5): PMA/510(K) # CHANGED FROM UNAVAILABLE TO K203768.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL REACHED 356 MG/DL. THE POD WAS WORN BETWEEN 4 AND 24 HOURS ON THE ABDOMEN. IT WAS NOTED THAT THE ADHESIVE WAS NOT SECURE AND THE CANNULA DISLODGED FROM THE SITE. HYPERGLYCEMIA TREATED WITH A NEW POD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1718132 OMNIPOD 5 POD ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP QFG INSULET CORPORATION PT-000435 PP1K02072231
991174 OMNIPOD 5 POD ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP QFG INSULET CORPORATION PT-000435 PP1K02072231

Patients

Seq Age Sex Outcome Treatment
1 16 YR Male