FDA Adverse Event Malfunction Summary report: N

REMSTAR AUTO A-FLEX

MDR report key: 15043359 · Received July 18, 2022

Report

Report Number
2518422-2022-55660
Event Type
Malfunction
Date Received
July 18, 2022
Date of Event
March 7, 2022
Report Date
July 3, 2025
Manufacturer
RESPIRONICS, INC.
Product Code
BZD
UDI-DI
00606959005884
PMA / PMN Number
K131982
Removal / Correction Number
RES 88058
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

SECTION G4 (PMA/510K) HAS BEEN CORRECTED FROM K113068 TO K131982 IN THIS REPORT.

Additional Manufacturer Narrative · 0

THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER PREVIOUSLY RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. REPEATED ATTEMPTS TO HAVE THE DEVICE AND COMPONENTS RETURNED FOR EVALUATION AND INVESTIGATION WERE UNSUCCESSFUL. THE MANUFACTURER BELIEVES THEY WILL BE UNABLE TO GATHER ADDITIONAL INFORMATION. THE MANUFACTURER IS SUBMITTING A FINAL REPORT AT THIS TIME. IF PERTINENT INFORMATION BECOMES AVAILABLE TO THE MANUFACTURER AT A LATER DATE, AN ADDENDUM TO THIS FINAL REPORT WILL BE FILED. SECTION H6 UPDATED IN THIS REPORT.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION WAS RECEIVED ON JUN 28, 2025 AND SECTION H11 SHOULD BE REPORTED AS: THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE MANUFACTURER PREVIOUSLY RECEIVED INFORMATION ALLEGING VISUALIZATION OF PARTICLES RELATED TO A CPAP DEVICE'S SOUND ABATEMENT FOAM. THERE WAS NO REPORT OF PATIENT HARM OR INJURY. THE DEVICE WAS RETURNED TO THE THIRD PARTY SERVICE CENTER FOR FURTHER EVALUATION. THE DEVICE WAS EVALUATED. THERE WAS NO MENTION OF VISUAL FINDINGS TO THE EXTERNAL PART OF THE DEVICE. THE INTERNAL ASPECT OF THE DEVICE WAS INSPECTED. THE DEVICE POWERED ON AND AIRFLOW WAS CONFIRMED. THE DEVICE'S DOWNLOADED LOGS WERE REVIEWED. THERE WERE NO ERRORS FOUND. THE THIRD PARTY SERVICE CENTER CONCLUDES THAT THEY COULD NOT CONFIRM THE CUSTOMER'S ALLEGATION AND THERE WAS NO VISIBLE FOAM DEGRADATION. DEVICE WAS SCRAPPED. SECTIONS D8, D9, H2, H3 AND H6 HAS BEEN UPDATED IN THIS REPORT.

Description of Event or Problem · 0

THE MANUFACTURER RECEIVED INFORMATION ALLEGING AN ISSUE RELATED TO CPAP DEVICE'S SOUND ABATEMENT FOAM. THE PATIENT ALLEGED VISUALIZATION OF PARTICLES. THERE WAS NO ALLEGATION OF SERIOUS OR PERMANENT HARM OR INJURY. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2873352 REMSTAR AUTO A-FLEX VENTILATOR, NON-CONTINUOUS (RESPIRATOR) BZD RESPIRONICS, INC. DS560S 00606959005884

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown