640G INSULIN PUMP MMT-1711P
Report
- Report Number
- 2032227-2022-288974
- Event Type
- Malfunction
- Date Received
- July 18, 2022
- Date of Event
- April 18, 2016
- Report Date
- July 17, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- Removal / Correction Number
- Z-0958-2020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
"THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. CORRECTION/REMOVAL NUMBER: Z-0958-2020. (B)(4). THIS RECORD WAS CONVERTED. FOR ADDITIONAL INFORMATION ABOUT THE ANALYSIS, PLEASE SEE SAP. UNIT RECEIVED WITH MINOR SCRATCHED LCD WINDOW, CRACKED KEYPAD AT SELECT BUTTON, CRACKED CASE NEAR BELT CLIP RAILS CORNERS AND CRACKED RETAINER. OZ (B)(6) 2017. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ""DEFECTS"" OR HAS ""MALFUNCTIONED"". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT."
MEDTRONIC RECEIVED INFORMATION THAT DAMAGE (PHYSICAL OR COSMETIC) OCCURRED. THERE WAS NO ADVERSE IMPACT OR CONSEQUENCE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 413138 | 640G INSULIN PUMP MMT-1711P | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1711P |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |