FDA Adverse Event Death Summary report: N

PUMP MMT-1780KPK 670G PATHWAY BLACK MG

MDR report key: 15038446 · Received July 18, 2022

Report

Report Number
2032227-2022-290151
Event Type
Death
Date Received
July 18, 2022
Date of Event
May 18, 2019
Report Date
July 18, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
00643169939219
Removal / Correction Number
Z-0956-2020
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

"THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. (B)(4). S/W 3.18E. THE UNIT DID NOT HAVE A BATTERY INSTALLED WHEN RECEIVED. UNIT PASSED THE FUNCTIONAL TEST, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND THE DAT TEST AT 0.08735 INCHES. UNIT UPLOADED PROPERLY USING CARELINK. UNIT HAD SCRATCHED CASE, CRACKED BATTERY TUBE THREADS, CRACKED CASE AT THE CORNER OF THE BELT CLIP RAIL, PILLOWING KEYPAD OVERLAY AND CRACKED RETAINER THE TEST P-CAP AND RESERVOIR DOES LOCK IN PLACE IN THE RESERVOIR COMPARTMENT. DATA ANALYSIS: THE FORMATTED HISTORY FILE LISTS DATA FROM 11/12/2018 TO 02/05/2019. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ""DEFECTS"" OR HAS ""MALFUNCTIONED"". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT."

Description of Event or Problem · 0

IT WAS REPORTED VIA PHONE CALL THAT THE CUSTOMER PASSED AWAY IN HOSPITAL. THE CUSTOMER WAS HOSPITALIZED ON (B)(6) 2019 DUE TO DIABETES. THE CAUSE OF DEATH WAS COMPLICATION OF DIABETES. THE CALLER STATED THAT THE CUSTOMER HAD DIABETES THAT MAY HAVE LED TO THE CUSTOMER'S PASSING. THE CUSTOMER REPORTED FUNGAL MENINGITIS WHICH MAY HAVE CONTRIBUTED OR LED UP TO PASSING. THE CALLER DECLINED TO RETURN THE INSULIN PUMP FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
959824 PUMP MMT-1780KPK 670G PATHWAY BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KPK HG1UTKB 00643169939219

Patients

Seq Age Sex Outcome Treatment
1 64 YR Unknown Death