BD KIESTRA INOQULA TLA
Report
- Report Number
- 3010141591-2022-00011
- Event Type
- Injury
- Date Received
- July 15, 2022
- Date of Event
- June 16, 2022
- Report Date
- August 5, 2022
- Manufacturer
- KIESTRA LAB AUTOMATION
- Product Code
- JTC
- UDI-DI
- 00382904472139
- PMA / PMN Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
COMMON DEVICE NAME: EQUIPMENT, LABORATORY, GENERAL PURPOSE, LABELED AND PROMOTED FOR A SPECIFIC MEDICAL USE. MEDICAL DEVICE EXPIRATION DATE: NA. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H 6. INVESTIGATION SUMMARY: THE FIELD SERVICE ENGINEER REPORTED THAT THE COVER OF THE BARCODA MODULE ON THE INSTRUMENT BD KIESTRA INOQULA TLA (MATERIAL #447213, SERIAL NUMBER (B)(6)) FELL DOWN ON LAB TECHNICIANS HAND CAUSING AN INJURY. ISSUE WAS CONFIRMED BY BD. THE FIELD SERVICE ENGINEER REPLACED THE TWO PNEUMATIC CYLINDERS ON THE BARCODA COVER TO ASSURE PROPER HOLD. THE INSTRUMENT WAS LEFT FULLY OPERATIONAL. BD HAS ISSUED A CORRECTIVE AND PREVENTIVE ACTION (CAPA) TO INVESTIGATE THE ISSUE FURTHER. DESIGN HISTORY RECORD (DHR) REVIEW IS NOT REQUIRED FOR THIS COMPLAINT. THE COMPLAINT WAS EVALUATED VIA OTHER ELEMENTS OF THE INVESTIGATION.
IT WAS REPORTED THAT THE BD KIESTRA INOQULA TLA- BARCODA LID CAME DOWN ON TECH FINGER. TECHNICIAN REPORTS THAT A LAB TECH HAD THE BARCODA LID COME DOWN ON THEIR FINGER. THE TECH WENT TO THE EMERGENCY ROOM AND HAD X-RAYS. NO BROKEN BONE FOUND. NO BLOOD REPORTED, TECH HAS A SMALL BRUISE, A BAND-AID WAS GIVEN.
IT WAS REPORTED THAT THE BD KIESTRA INOQULA TLA- BARCODA LID CAME DOWN ON TECH FINGER TECHNICIAN REPORTS THAT A LAB TECH HAD THE BARCODA LID COME DOWN ON THEIR FINGER. THE TECH WENT TO THE EMERGENCY ROOM AND HAD X-RAYS. NO BROKEN BONE FOUND. NO BLOOD REPORTED, TECH HAS A SMALL BRUISE, A BAND-AID WAS GIVEN.