FDA Adverse Event Malfunction Summary report: N

ENDOFLATOR 50, W/INTEGRATED SCB MODULE

MDR report key: 15031354 · Received July 15, 2022

Report

Report Number
9610617-2022-00123
Event Type
Malfunction
Date Received
July 15, 2022
Report Date
November 23, 2022
Manufacturer
KARL STORZ SE & CO. KG
Product Code
FCX
UDI-DI
04048551326657
PMA / PMN Number
K161554
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE EVALUATION CONFIRMED A PRESSURE SENSOR INTERRUPTION ERROR OCCURRED. ACCORDING TO THE TREND REVIEW, ONE PREVIOUS CASE CAUSED A SERIOUS INJURY. THUS, A REPORT IS REQUIRED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED AGAIN FOR EVALUATION, AS IT SUBSEQUENTLY SHUT OFF DURING A CASE (REF. MDR 9610617-2022-00132) AND, PER OUR US INSPECTION TEAM, THE UNIT POWERED OFF DURING ENDURANCE TEST. THE UNIT WAS FORWARDED TO THE MANUFACTURER IN GERMANY FOR FURTHER INVESTIGATION. AS PER THE MANUFACTURER'S INVESTIGATION REPORT: THE PRODUCT INSPECTION CANNOT BE PERFORMED AT KST BECAUSE THE DEVICE HAS BEEN LOST INTERNALLY. BASED ON THE INVESTIGATION RESULTS OF THE US COLLEAGUES, THE DEVICE FAILED DURING THE LONG-TERM TEST. THE MOST PROBABLE ROOT CAUSE IS A FAULT WITH THE POWER SUPPLY. THIS COULD BE A PROBLEM WITH THE POWER INPUT SOCKET OR THE FUSE, AMONG OTHER THINGS. IF THE DEVICE COULD BE FOUND, THE REPORT WILL BE RE-OPENED AND ADJUSTED ACCORDINGLY. A WARNING IN THE IFU IS LISTED TO ALWAYS KEEP A REPLACEMENT DEVICE ON HAND IN CASE THE DEVICE SHOULD FAIL.

Description of Event or Problem · 0

CUSTOMER REPORTED THAT THE ENDOFLATOR 50 UNIT STOPPED PUSHING GAS IS TWO SEPARATE CASES. THERE WAS NO HARM TO THE PATIENT. THE DATE OF THE EVENTS COULD NOT BE CONFIRMED. THIS REPORT REPRESENTS CASE 2 OF 2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2756364 ENDOFLATOR 50, W/INTEGRATED SCB MODULE LAPAROSCOPIC GAS DISTENSION SYSTEM FCX KARL STORZ SE & CO. KG UI500 04048551326657

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other