FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE

MDR report key: 15019166 · Received July 14, 2022

Report

Report Number
1024879-2022-00363
Event Type
Malfunction
Date Received
July 14, 2022
Date of Event
July 8, 2022
Report Date
July 27, 2022
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
FMI
UDI-DI
50382903686075
PMA / PMN Number
K982541
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVAL: YES. D9: RETURNED TO MANUFACTURER ON: 2022-07-27. H6: INVESTIGATION SUMMARY: BD RECEIVED 48 SAMPLES AND 2 PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR DEFECTIVE LOCKING MECHANISM WAS OBSERVED. ADDITIONALLY, 20 OF THE CUSTOMER SAMPLES WERE EVALUATED BY FUNCTIONAL TESTING AND THE INDICATED FAILURE MODE FOR DEFECTIVE LOCKING MECHANISM WITH THE INCIDENT LOT WAS NOT OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE DEFECTIVE LOCKING MECHANISM. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS."

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THE SAFETY SHIELD WAS SLIDING OFF OR BREAKING OFF DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS SAFETY SHIELD SLIDING OFF OR BREAKING OFF DEVICE FOR PRODUCT 368607 LOT 2090159.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE THE SAFETY SHIELD WAS SLIDING OFF OR BREAKING OFF DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS SAFETY SHIELD SLIDING OFF OR BREAKING OFF DEVICE FOR PRODUCT 368607; LOT 2090159.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2284801 BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON, DICKINSON & CO., (BD) 368607 2090159 50382903686075

Patients

Seq Age Sex Outcome Treatment
1 Unknown