FDA Adverse Event Malfunction Summary report: N

BD DISCARDIT¿ II SYRINGE

MDR report key: 15018406 · Received July 14, 2022

Report

Report Number
3002682307-2022-00182
Event Type
Malfunction
Date Received
July 14, 2022
Date of Event
June 15, 2022
Report Date
September 28, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 31AUG2022. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309110 AND LOT NUMBER 2201185. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. TO AID IN THE INVESTIGATION OF THIS ISSUE, THE AFFECTED SAMPLE WAS RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH ANALYSIS OF THE SAMPLE, LEAKAGE WAS OBSERVED, CONFIRMING THE REPORTED DEFECT. IT HAS BEEN CONCLUDED THAT THIS DEFECT MOST LIKELY RESULTED FROM DAMAGE TO THE PLUNGER LIP COMPONENT. THIS TYPE OF DAMAGE MAY RESULT FROM THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING FACILITY OR WITHIN THE PLUNGER ASSEMBLY MACHINE. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD DISCARDIT¿ II SYRINGE LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: ¿SYRINGE IS LEAKING."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD DISCARDIT¿ II SYRINGE LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM GERMAN: ¿SYRINGE IS LEAKING.".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
426525 BD DISCARDIT¿ II SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2201185

Patients

Seq Age Sex Outcome Treatment
1 Unknown