FDA Adverse Event Other Summary report: N

GELWEAVE SIENA COLLARED 4 BRANCH PLEXUS RADIOPAQUE MARKERS

MDR report key: 1501501 · Received October 8, 2009

Report

Report Number
9612515-2009-00009
Event Type
Other
Date Received
October 8, 2009
Date of Event
June 16, 2009
Report Date
October 6, 2009
Manufacturer
VASCUTEK LTD.
Product Code
DSY
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

METHOD: QC AND MANUFACTURING RECORDS WERE REVIEWED. AN OFF-CUT OF THE GRAFT HAS BEEN RETURNED TO VASCUTEK FOR ANALYSIS. RESULTS CODE: THERE WERE TOTAL OF THREE UNITS IN THE BATCH. ALL GRAFTS WERE GELATIN SEALED. AFTER, SEALING, 100% OF ALL GRAFTS ARE WHOLE GRAFT POROSITY TESTED. THIS INVOLVES PRESSURIZING THE DEVICE WITH AN ISOPROPANOL ALCOHOL / GLYCEROL MIXTURE AND MEASURING THE FLOW RATE. THE MAXIMUM PERMISSIBLE FLOW RATE WAS DETERMINED BY COMPARISON WITH CITRATED COWS BLOOD. THE QUALITY CONTROL AND MANUFACTURING RECORDS HAVE BEEN REVIEWED AND THERE IS NOTHING TO SUGGEST ANY PROBLEMS. THE LEAK TEST RESULTS WERE WITHIN OUR SPECIFIED QUALITY CONTROL LIMITS FOR THIS PRODUCT. ALL WERE DISTRIBUTED IN OCTOBER 2007. NO OTHER REPORTS HAVE BEEN RECEIVED FOR THE OTHER TWO UNITS. ANALYSIS OF THE GRAFT IS TO TAKE PLACE. CONCLUSION CODE: NO CONCLUSION CAN BE IDENTIFIED AT THIS TIME.

Description of Event or Problem · 1

THE GELWEAVE SIENA COLLARED 4 BRANCH PLEXUS WITH RADIOPAQUE MARKERS IS CE MARKED. THE EVENT OCCURRED IN A HOSPITAL IN (B)(6). THE EVENT IS BEING REPORTED DUE TO PROLONGATION OF THE OPERATION. THE PATIENT ALSO HAD A POST OPERATIVE BRAIN SCAN. THERE WERE TWO AREAS OF CEREBRAL HEMORRHAGE. THE PATIENT UNDERWENT A SURGICAL ARCH REPLACEMENT. AFTER HEPARIN REVERSAL, THERE WAS BLOOD LEAKAGE FROM THE PROSTHESIS, WHICH RESULTED IN PROLONGATION. THE PATIENT RECEIVED BLOOD REPLACEMENT, PLATELETS, RED BLOOD CELLS AND FRESH FROZEN PLASMA. THE PATIENT'S COAGULATION FUNCTION DETERIORATED, REQUIRING ADDITIONAL PLATELETS, FRESH FROZEN PLASMA AND LOCAL ADMINISTRATION OF CELOX, WHICH CONTROLS ARTERIAL BLEEDING. THE CAUSE OF THE CEREBRAL HEMORRHAGE IS NOT GIVEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GELWEAVE SIENA COLLARED 4 BRANCH PLEXUS RADIOPAQUE MARKERS VASCULAR PROSTHESIS DSY VASCUTEK LTD. UNK 82648 0465

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention