IMAGE1 S CONNECT
Report
- Report Number
- 15013353
- Event Type
- Malfunction
- Date Received
- July 14, 2022
- Date of Event
- May 6, 2021
- Report Date
- June 30, 2022
- Manufacturer
- KARL STORZ ENDOSCOPY-AMERICA, INC.
- Product Code
- GWG
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- DC
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE KARL STORZ ENDOSCOPY EQUIPMENT TC 201 AND TC 304 MODELS WERE HAVING ISSUES WHERE THE POWER WOULD NOT SHUT OFF AND THE UNITS OCCASIONALLY OVERHEATED. STORZ IDENTIFIED A FIX TO INSULATE THE FRONT COVER OF THE EQUIPMENT, ULTIMATELY REDUCING WATER INTRUSION. IT SEEMED AS THOUGH THE DESIGN OF THE DEVICE WAS NOT MADE TO HANDLE THE CLEANING FLUID, AND WHEN FLUID CAME INTO THE FRONT PANEL, THE POWER BUTTON MALFUNCTIONED. MANUFACTURER RESPONSE FOR ENDOSCOPY CAMERA CONTROL UNIT, IMAGE1 S CONNECT II (PER SITE REPORTER). THE MANUFACTURER DID A FULL EVALUATION, AND FOUND THAT ADDITIONAL INSULATION WAS NECESSARY TO REDUCE WATER INTRUSION ON THE POWER BUTTONS AND ELECTRICAL NETWORKS.
THE KARL STORZ ENDOSCOPY EQUIPMENT TC 201 AND TC 304 MODELS WERE HAVING ISSUES WHERE THE POWER WOULD NOT SHUT OFF AND THE UNITS OCCASIONALLY OVERHEATED. STORZ IDENTIFIED A FIX TO INSULATE THE FRONT COVER OF THE EQUIPMENT, ULTIMATELY REDUCING WATER INTRUSION. IT SEEMED AS THOUGH THE DESIGN OF THE DEVICE WAS NOT MADE TO HANDLE THE CLEANING FLUID, AND WHEN FLUID CAME INTO THE FRONT PANEL, THE POWER BUTTON MALFUNCTIONED. MANUFACTURER RESPONSE FOR ENDOSCOPY CAMERA CONTROL UNIT, IMAGE1 S CONNECT II (PER SITE REPORTER): THE MANUFACTURER DID A FULL EVALUATION, AND FOUND THAT ADDITIONAL INSULATION WAS NECESSARY TO REDUCE WATER INTRUSION ON THE POWER BUTTONS AND ELECTRICAL NETWORKS.
THE KARL STORZ ENDOSCOPY EQUIPMENT TC 201 AND TC 304 MODELS WERE HAVING ISSUES WHERE THE POWER WOULD NOT SHUT OFF AND THE UNITS OCCASIONALLY OVERHEATED. STORZ IDENTIFIED A FIX TO INSULATE THE FRONT COVER OF THE EQUIPMENT, ULTIMATELY REDUCING WATER INTRUSION. IT SEEMED AS THOUGH THE DESIGN OF THE DEVICE WAS NOT MADE TO HANDLE THE CLEANING FLUID, AND WHEN FLUID CAME INTO THE FRONT PANEL, THE POWER BUTTON MALFUNCTIONED. MANUFACTURER RESPONSE FOR ENDOSCOPY CAMERA CONTROL UNIT, IMAGE1 S CONNECT II (PER SITE REPORTER): THE MANUFACTURER DID A FULL EVALUATION, AND FOUND THAT ADDITIONAL INSULATION WAS NECESSARY TO REDUCE WATER INTRUSION ON THE POWER BUTTONS AND ELECTRICAL NETWORKS.
THE KARL STORZ ENDOSCOPY EQUIPMENT TC 201 AND TC 304 MODELS WERE HAVING ISSUES WHERE THE POWER WOULD NOT SHUT OFF AND THE UNITS OCCASIONALLY OVERHEATED. STORZ IDENTIFIED A FIX TO INSULATE THE FRONT COVER OF THE EQUIPMENT, ULTIMATELY REDUCING WATER INTRUSION. IT SEEMED AS THOUGH THE DESIGN OF THE DEVICE WAS NOT MADE TO HANDLE THE CLEANING FLUID, AND WHEN FLUID CAME INTO THE FRONT PANEL, THE POWER BUTTON MALFUNCTIONED. MANUFACTURER RESPONSE FOR ENDOSCOPY CAMERA CONTROL UNIT, IMAGE1 S CONNECT II (PER SITE REPORTER): THE MANUFACTURER DID A FULL EVALUATION, AND FOUND THAT ADDITIONAL INSULATION WAS NECESSARY TO REDUCE WATER INTRUSION ON THE POWER BUTTONS AND ELECTRICAL NETWORKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 464060 | IMAGE1 S CONNECT | ENDOSCOPE, NEUROLOGICAL | GWG | KARL STORZ ENDOSCOPY-AMERICA, INC. | TC201US |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |