THERMOCOOL SF CARTOXP,B-D,TC
Report
- Report Number
- 2029046-2022-01602
- Event Type
- Injury
- Date Received
- July 13, 2022
- Date of Event
- June 15, 2022
- Report Date
- August 29, 2022
- Manufacturer
- BIOSENSE WEBSTER INC
- Product Code
- LPB
- UDI-DI
- 10846835003154
- PMA / PMN Number
- P030031/S025
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE BIOSENSE WEBSTER, INC. PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETED AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL 3500A REPORT WILL BE SUBMITTED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. BIOSENSE WEBSTER MANUFACTURER'S REFERENCE NUMBER (B)(4) HAS TWO REPORTS: (1) MFR # 2029046-2022-01602 FOR PRODUCT CODE BNI35BDH (THERMOCOOL® SF NAV BI-DIRECTIONAL CATHETER) (2) MFR # 2029046-2022-01603 FOR PRODUCT CODE BNI35BDH (THERMOCOOL® SF NAV BI-DIRECTIONAL CATHETER).
ADDITIONAL INFORMATION WAS RECEIVED ON 01-AUG-2022. THE FIRST ABLATION CATHETER¿S LOT NUMBER WAS 30703467L, AND THE SECOND WAS 30695591L. BOTH CATHETERS HAD HIGH IMPEDANCE ISSUE. NO ISSUE OCCURRED ON THE SMARTABLATE GENERATOR. TRANSSEPTAL PUNCTURE WAS CONDUCTED WITH RF NEEDLE. BASED ON THIS ADDITIONAL INFORMATION, THE CONCOMITANT PRODUCT SECTION WAS UPDATED. CORRECT CATHETER SETTINGS WERE SELECTED ON THE GENERATOR. NO ISSUE OBSERVED WITH PUMP. THE PUMP WAS SWITCHING FROM ¿LOW¿ TO ¿HIGH¿ FLOW DURING ABLATION. NO ERROR MESSAGE OBSERVED DURING THE PROCEDURE. DF, REALTIME GRAPH, VECTOR FORCE VISUALIZATION FEATURES WERE USED. ADDITIONAL INFORMATION WAS RECEIVED ON 8-AUG-2022. IT WAS REPORTED THAT THERE WAS NO CHAR /COAGULUM/THROMBUS/CLOT WAS CONFIRMED. THE PATIENT WAS ANTICOAGULATED, BUT ACT PRACTICE WAS UNKNOWN. THE PATIENT HAS NOT EXHIBITED ANY NEUROLOGICAL SYMPTOMS SINCE THE PROCEDURE WAS COMPLETED. HEPARINIZED NORMAL SALINE WAS USED AS THE IRRIGATION FLUID. VISITAG MODULE WAS USED WITH THE RESPIRATION FILTER. THE DEVICE EVALUATION WAS COMPLETED ON 18-AUG-2022. THE PRODUCT WAS RETURNED TO BIOSENSE WEBSTER (BWI) FOR EVALUATION. A VISUAL INSPECTION AND AN EVALUATION OF ALL FEATURES OF THE DEVICE WERE PERFORMED FOLLOWING BWI PROCEDURES. VISUAL ANALYSIS REVEALED NO DAMAGE OR ANOMALIES ON THE DEVICE. PER THE EVENT, SEVERAL TESTS WERE PERFORMED. THE MAGNETIC, ELECTRICAL, TEMPERATURE, AND FORCE FEATURES WERE TESTED, AND NO ISSUES WERE OBSERVED. IN ADDITION, THE PRODUCT WAS DEFLECTING CORRECTLY. THE DEVICE WAS PARTIALLY IRRIGATED; THE DEVICE WAS DISSECTED, AND THE DISTAL HOLES OF THE DOME WERE FOUND OBSTRUCTED WITH REDDISH MATERIAL INSIDE. THE ISSUE WITH THE DISTAL HOLES OF THE DOME BEING OCCLUDED IS CONSIDERED NOT MDR REPORTABLE. THE MOST LIKELY CONSEQUENCE IS AN INTRAPROCEDURAL DELAY. THE POTENTIAL RISK THAT IT COULD CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR DEATH IS REMOTE. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR LOT 30703467L, AND NO INTERNAL ACTIONS RELATED TO THE COMPLAINT WERE FOUND DURING THE REVIEW. THE ROOT CAUSE OF THE ADVERSE EVENT REMAINS UNKNOWN; HOWEVER, NO DEVICE MALFUNCTION WAS FOUND THAT COULD HAVE CAUSED THIS. THE INSTRUCTION FOR USE (IFU) STATES THAT CAREFUL CATHETER MANIPULATION MUST BE PERFORMED TO AVOID CARDIAC DAMAGE, PERFORATION, OR TAMPONADE. IF THE RADIOFREQUENCY (RF) GENERATOR DOES NOT DISPLAY TEMPERATURE, VERIFY THAT THE APPROPRIATE CABLE IS PLUGGED INTO THE RF GENERATOR. AS PART OF BIOSENSE WEBSTER¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. BIOSENSE WEBSTER MANUFACTURER'S REFERENCE NUMBER PC-001143942 HAS TWO REPORTS: (1) MFR # 2029046-2022-01602 FOR PRODUCT CODE BNI35BDH (THERMOCOOL® SF NAV BI-DIRECTIONAL CATHETER) (2) MFR # 2029046-2022-01603 FOR PRODUCT CODE BNI35BDH (THERMOCOOL® SF NAV BI-DIRECTIONAL CATHETER) ON 01-AUG-2022, ADDITIONAL INFORMATION WAS RECEIVED AND THE CODE OF "CHAR" (IMDRF CODE OF DEVICE CONTAMINATION WITH BODY FLUID (A180103)) HAS BEEN REMOVED SINCE IT WAS NOTED THAT THE PICTURE THAT WAS INITIALLY PRESENTED WAS FROM ANOTHER ISSUE NOT RELATED TO THIS COMPLAINT. THEREFORE, H6. MEDICAL DEVICE PROBLEM CODE REFLECTS THE TWO CODES THAT REMAIN APPLICABLE.
IT WAS REPORTED THAT A MALE PATIENT UNDERWENT A PREMATURE VENTRICULAR CONTRACTION (PVC) ABLATION PROCEDURE WITH TWO THERMOCOOL® SF NAV BI-DIRECTIONAL CATHETERS (2) AND SUFFERED CARDIAC TAMPONADE REQUIRING PERICARDIOCENTESIS. DURING THE PVC PROCEDURE, THE CATHETER WAS ADVANCED TO THE TARGET SITE RIGHT VENTRICLE (RV), IMPEDANCE WAS SHOWN AS ABOUT 200 OHM. THEY EXCHANGED TO A NEW CABLE BUT THERE IS NO CHANGE ON THE IMPEDANCE, AND THEN, THE CATHETER WAS REPLACED WITH A NEW CATHETER OF THE SAME TYPE BUT THE VALUE DID NOT CHANGE. THEREFORE, WHEN THE CATHETER WAS MOVED TO ANOTHER POSITION, THE VALUE WAS AROUND 130 ~ 180 OHM. FIRST ABLATION WAS CONDUCTED AS 30W/5SEC AND IMPEDANCE WAS 53 OHM DOWN FROM THE STARTING OF 234 OHM. THE SECOND ABLATION WAS CONDUCTED AT 30W/5SEC AND IMPEDANCE WAS 52 DOWN FROM THE STARTING OF 204 OHM. AFTER SWITCHING TO LIFELINE ABLAZE, DECREASED BLOOD PRESSURE WAS CONFIRMED. THEN, SETTING WAS SET AT 50 W WITH ABLAZE (W WAS CHANGED AT 1 ~ 40 W) AND ABLATION WAS STARTED AT 135 OHM (MAX 152 OHM), 27 SECONDS ABLATION WAS PERFORMED. CARDIAC TAMPONADE WAS CONFIRMED BY ECHO, AND DRAINAGE WAS STARTED. ECHO WAS CONFIRMED AGAIN, BLOOD PRESSURE RETURNED TO NORMAL LIMITS, AND THE PATIENT RETURNED TO THE ROOM. THE PHYSICIAN'S OPINION ON THE RELATIONSHIP BETWEEN THE EVENT AND THE PRODUCT IS THAT THERE MAY HAVE BEEN AN IMPEDANCE ABNORMALITY. THE ADVERSE EVENT OCCURRED ON (B)(6) 2022. IT WAS DISCOVERED POST USE OF BIOSENSE WEBSTER CATHETERS. THE TWO DEFECTIVE BWI CATHETERS WERE REMOVED DUE TO PRODUCT MALFUNCTION, THE ABLATION PROCEDURE WAS CONTINUED WITH ANOTHER COMPANY'S CATHETER AND ADVERSE EVENT WAS CONFIRMED WHEN THE ABLATION WAS PERFORMED WITH THE OTHER COMPANY'S ABLATION CATHETER. THE PHYSICIAN¿S OPINION ON THE CAUSE OF THIS ADVERSE EVENT IS THAT IT WAS DUE TO THE PROCEDURE. THE PATIENT OUTCOME OF THE ADVERSE EVENT IS IMPROVED. PRIOR TO NOTING THE PERICARDIAL EFFUSION, ABLATION WAS PERFORMED. THERE WAS NO EVIDENCE OF A STEAM POP. THE EVENT OCCURRED DURING THE ABLATION PHASE. THE ABLATION NEVER CONTINUED BEYOND THE CUT-OFF VALUE. GENERATOR PARAMETERS WAS POWER CONTROL MODE. THE COMPLAINT PRODUCTS WILL BE RETURNED FOR ANALYSIS. NO ERROR WAS MESSAGE OBSERVED DURING THE PROCEDURE. NO ISSUE RELATED TO THE TEMPERATURE AND FLOW ON CATHETER WAS OBSERVED. A SMARTABLATE GENERATOR WITH PRODUCT CODE M4900207 AND SERIAL NUMBER WAS (B)(4) WAS USED. THE CATHETER WILL BE RETURNED FOR ANALYSIS. NO NEUROLOGICAL SYMPTOMS WERE OBSERVED ON THE PATIENT. THE DURATION OF ABLATION USED WAS LESS THAN 60 SECONDS. THIS COMPLAINT WILL BE CODED WITH CHAR. FOLLOW UP HAS BEEN DONE TO CLARIFY THE RESPONSE RECEIVED THAT NO CHAR /COAGULUM/THROMBUS/CLOT WAS CONFIRMED AS THERE IS DISCREPANCY WITH THE PHOTOGRAPH PROVIDED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS COMPLAINT WILL BE REASSESSED. SINCE THE EVENT (CARDIAC TAMPONADE) IS LIFE THREATENING AND REQUIRED INTERVENTION AND PROLONGED HOSPITALIZATION TO PREVENT PERMANENT IMPAIRMENT OF A BODY FUNCTION OR PERMANENT DAMAGE TO A BODY STRUCTURE, IS TO BE CONSIDERED SERIOUS AND MDR-REPORTABLE. THE HIGH IMPEDANCE WAS ASSESSED AS NOT MDR REPORTABLE. SINCE THE USER-DEFINED CUT-OFF WAS NOT EXCEEDED THE POTENTIAL THAT IT COULD CAUSE OR CONTRIBUTE TO A DEATH, SERIOUS INJURY, OR OTHER SIGNIFICANT ADVERSE EVENT, IS REMOTE. CHAR WAS ASSESSED AS NOT MDR REPORTABLE. CHAR IS A PHYSICAL PHENOMENON OF RADIOFREQUENCY (RF) ENERGY DELIVERY AND CAN BE THE NORMAL RESULT OF THE ABLATION PROCESS. THE PRESENCE OF CHAR ON THE ELECTRODES DOES NOT REPRESENT A SERIOUS INJURY IN ITSELF, NOR IS IT NECESSARILY THE RESULT OF DEVICE MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2285440 | THERMOCOOL SF CARTOXP,B-D,TC | CARDIAC ABLATION PERCUTANEOUS CATHETER | LPB | BIOSENSE WEBSTER INC | BNI35BDH | 30703467L | 10846835003154 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male | Life Threatening| R | SMARTABLATE GEN. KIT (JAPAN)| SMARTABLATE GEN. KIT (JAPAN)| THERMOCOOL SF CARTOXP,B-D,TC| UNK BRAND CABLE| UNK BRAND CABLE| UNK BRAND RADIOFREQUENCY (RF) NEEDLE| UNK JAPAN LIFELINE ABLAZE CATHETER |