FDA Adverse Event Malfunction Summary report: N

BD INTIMA II¿ IV CATHETER PRN ADAPTER

MDR report key: 15008703 · Received July 13, 2022

Report

Report Number
3014704491-2022-00274
Event Type
Malfunction
Date Received
July 13, 2022
Date of Event
June 13, 2022
Report Date
July 1, 2022
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: IN RESPONSE TO THE EVENT REPORTED, A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 2011039. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. OUR ENGINEERS WERE ABLE TO DETERMINE THAT THE MOST LIKELY ROOT CAUSE FOR THIS EVENT IS THE FORCING OF FLUID THROUGH THE DEVICE DURING INJECTION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INTIMA II¿ IV CATHETER PRN ADAPTER BURST DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE: "RADIOLOGISTS FOUND THAT PART OF THE INDWELLING NEEDLE INFUSION SET WAS INSERTED AND BURST WHEN THEY WERE USING THE CLOSED VENOUS INDWELLING NEEDLE FOR INFUSION OF PATIENTS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470264 BD INTIMA II¿ IV CATHETER PRN ADAPTER INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 2011039

Patients

Seq Age Sex Outcome Treatment
1 Unknown