FDA Adverse Event Malfunction Summary report: N

BD DISCARDIT¿ II SYRINGE

MDR report key: 15008410 · Received July 13, 2022

Report

Report Number
3002682307-2022-00180
Event Type
Malfunction
Date Received
July 13, 2022
Date of Event
March 20, 2022
Report Date
August 1, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
UDI-DI
00382903096572
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 309110 AND LOT NUMBER 2110203. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR RETURN, RETAINED SAMPLES WERE OBTAINED FOR EVALUATION; HOWEVER, THE RETAINED SAMPLES DISPLAYED NO SIGNS OF LEAKAGE. WITH THE LIMITED INVESTIGATION RESULTS, AN EXACT CAUSE COULD NOT BE DETERMINED FOR THIS INCIDENT. HOWEVER, BASED ON THE CUSTOMER FEEDBACK, IT IS POSSIBLE THAT THIS INCIDENT RESULTED FROM DAMAGE TO THE PLUNGER LIP COMPONENT. THIS TYPE OF DAMAGE MAY RESULT FROM THE HANDLING OF THE PRODUCT THROUGH THE MANUFACTURING FACILITY OR WITHIN THE PLUNGER ASSEMBLY MACHINE. BASED ON THE PREVENTIVE MEASURES IN PLACE, WE BELIEVE THIS WAS AN ISOLATED INCIDENT WITH AN UNLIKELY CHANCE OF RECURRENCE. OUR QUALITY TEAM WILL CONTINUE TO CLOSELY MONITOR THE MANUFACTURING PROCESS FOR SIGNS OF THIS POTENTIAL DEFECT AND ANY EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD DISCARDIT¿ II SYRINGE EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: "LIQUID LEAKAGE BETWEEN THE SYRINGE BARREL AND THE PLUNGER."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD DISCARDIT¿ II SYRINGE EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM FRENCH: "LIQUID LEAKAGE BETWEEN THE SYRINGE BARREL AND THE PLUNGER."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
417961 BD DISCARDIT¿ II SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2110203 00382903096572

Patients

Seq Age Sex Outcome Treatment
1 Unknown