PLUM A+/A+3 (V 11.X & 13.X)
Report
- Report Number
- 9615050-2022-00142
- Event Type
- Malfunction
- Date Received
- July 13, 2022
- Date of Event
- June 1, 2022
- Report Date
- June 17, 2022
- Manufacturer
- ICU MEDICAL COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K070398
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- 003
Narratives
THE DEVICE HAS BEEN RECEIVED. AN INVESTIGATION IS PENDING.
THE DEVICE WAS IN GOOD GENERAL CONDITION. THE EVENT/ALARM LOGS WERE DOWNLOADED AND SAVED ON FILE. LOGS ATTACHED TO SR# (B)(6). THE LAST ENTRIES IN THE LOGS WERE ON THE (B)(6)2022 AT 7.40. AT 07.40 THE DEVICE IS PROGRAMMED AS FOLLOWS: (B)(6) 2022 07:40 LINEA A: DURATA: 00:56 (B)(6) 2022 07:40 LINEA A: VELOCITÀ:250, VDI:235,0 THIS PROGRAM COMPLETES SUCCESSFULLY WITHOUT ISSUE 56 MIN LATER AT 08.36. (B)(6) 2022 08:36 N161 VDI LINEA A COMPLETE THEN AT 08.41, THE DEVICE VTBI IS CHANGED TO ADD ANOTHER 30ML. (B)(6) 2022 08:41 LINEA A: DURATA: 00:07 (B)(6) 2022 08:41 LINEA A: VELOCITÀ:250, VDI:30,0 THIS PROGRAM COMPLETES SUCCESSFULLY WITHOUT ISSUE 7 MIN LATER AT 08.48. (B)(6) 2022 08:48 ARRESTO INFUSIONE: A, VOLUME INFUSO=265,0 (B)(6) 2022 08:48 PROGRAMMA LINEA A: COMPLETATO THIS SHOWS THAT THE DEVICE PERFORMED AS PROGRAMMED AND DELIVERED A TOTAL OF 265ML AS EXPECTED. THE DEVICE COMPLETED FULL PERFORMANCE VERIFICATION TESTING (PVT) WITHOUT ISSUE. THE CUSTOMER¿S PROTOCOL WAS PROGRAMMED AND COMPLETED SUCCESSFULLY DELIVERING A TOTAL OF 263.5ML. THE COMPLAINT IS NOT CONFIRMED, AND NO PROBABLE CAUSE COULD BE DETERMINED.
IT WAS REPORTED THAT, ON AN UNKNOWN DATE, THE PLUM A+ PUMP PRESENTS A MISMATCH BETWEEN THE LIQUID LIMITS SET AND THE LIQUIDS ADMINISTERED: THE INFUSION SET EMPTIES ABNORMALLY. THERE WAS NO PATIENT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1702128 | PLUM A+/A+3 (V 11.X & 13.X) | PUMP, INFUSION | FRN | ICU MEDICAL COSTA RICA LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | UNSPECIFIED LIQUID, UNK MFR. |