FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

MDR report key: 15001092 · Received July 13, 2022

Report

Report Number
8010047-2022-11824
Event Type
Malfunction
Date Received
July 13, 2022
Report Date
May 13, 2024
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FDS
UDI-DI
04953170305290
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CUSTOMER (B)(6) UNIV HEALTH ((B)(6) HEALTH) FACILITIES HAD CONCERNS CONNECTED WITH FINDINGS OF DEVICES INSPECTED BY BORESCOPE AND THE REPROCESSING AT THE VARIOUS (B)(6) HEALTH FACILITIES. ARISING FROM THE FINDINGS, DEVICES/EVENTS HAVE BEEN IDENTIFIED FOR THE (B)(6) HEALTH FACILITIES BELOW: (B)(6). THE ASSOCIATED DEVICES/EVENTS THAT ARE IDENTIFIED AT THIS TIME FOR INDIANA UNIV HEALTH MORGAN HOSPITAL INC. ARE AS BELOW: PATIENT IDENTIFIER: (B)(6), MODEL # CF-HQ190L, SERIAL NUMBER: (B)(4); (B)(6), PCF-H190L, (B)(4); (B)(6), GIF-H190, (B)(4); (B)(6), GIF-H190, (B)(4), (B)(6), GIF-H190, (B)(4), (B)(6), PCF-H180AL, (B)(4), (B)(6), GIF-H190, (B)(4), (B)(6), GIF-1TQ160, (B)(4). THIS MEDWATCH IS FOR PATIENT IDENTIFIER: (B)(6). IN-SERVICE WAS PERFORMED BY THE OLYMPUS ENDOSCOPY SUPPORT SPECIALIST (ESS) FOR THE FACILITY¿S REPROCESSING. THE CLEANING, DISINFECTION, AND STERILIZATION INFORMATION WAS REVIEWED DURING IN-SERVICE AND DOCUMENTED. BELOW ARE THE OBSERVATIONS MADE: DURING REPROCESSING IN-SERVICE IT WAS OBSERVED THAT THE CUSTOMER DOES NOT ALWAYS USE THE AIR/WATER CHANNEL CLEANING ADAPTER DURING THE PRE-CLEANING. THE CUSTOMER LEAK TESTS THE SCOPE FOR ONLY 10 SECONDS AND DOES NOT ANGULATE THE CONTROL KNOBS. CUSTOMER DOES NOT REMOVE THE SCOPE FROM THE WATER PRIOR TO REMOVING THE LEAK TESTER FROM THE SCOPE CONNECTOR. THE SCOPE CONNECTOR IS THE ONLY PART REMOVED FROM THE WATER. THE LEAK TESTER IS ONLY DISCONNECTED FROM THE ETHYLENE OXIDE (EO OR ETO) ETO VALVE ON THE SCOPE AND IS NOT TURNED OFF OR PULLED FROM THE MU-1 UNIT TO DEPRESSURIZE THE SCOPE. THE CUSTOMER USES THE CHANNEL OPENING CLEANING BRUSH PRIOR TO USING THE CHANNEL CLEANING BRUSH. THE BRUSHING IS DONE IN THE INCORRECT ORDER. THE CUSTOMER USES THIRD PARTY BRUSHES FROM MEDIVATORS. THE CUSTOMER ALSO DOES NOT WIPE THE SCOPE DOWN AFTER THE FLUSHING WITH DETERGENT SOLUTION AND DOES NOT LET IT SOAK FOR THE DETERGENT MANUFACTURER'S RECOMMENDED TIME. THE SCOPE DOES NOT SOAK AT ALL AND GOES DIRECTLY INTO THE RINSE SINK. THE CUSTOMER USES INTERCEPT DETERGENT. THE CUSTOMER USES THE SCOPE BUDDY PLUS FOR FLUSHING AND THE MEDIVATOR¿S DSD AUTOMATIC ENDOSCOPY REPROCESSOR (AER). THE CORRECT REPROCESSING METHOD PER THE INSTRUCTIONS FOR USE (IFU) WERE REVIEWED IN THE IN-SERVICE. CUSTOMER MANAGER WAS PROVIDED WITH A COPY OF THE DOCUMENTATION OF THE IN-SERVICE, POSTERS, AND WEBSITES FOR REPROCESSING VIDEOS. CUSTOMER WAS ADVISED TO FOLLOW ALL VALIDATED REPROCESSING METHODS LISTED IN THE REPROCESSING IFU. THE DEVICE IS NOT RETURNED, AS SUCH A DEFINITIVE ROOT CAUSE OF THE REPORTED COMPLAINT CANNOT BE DETERMINED AT THIS TIME.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION, TO UPDATE D8, AND TO CORRECT D9. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, THE ROOT CAUSE COULD NOT BE DETERMINED. IT WAS CONFIRMED THAT WHITE FLECKED SEDIMENT WAS FOUND AT THE BIOPSY CHANNEL PORT AND A THREAD LIKE FOREIGN MATERIAL WAS FOUND AT THE DISTAL END, HOWEVER, THE SPECIFIC MATERIALS COULD NOT BE IDENTIFIED. ADDITIONALLY, THE BRUSH WAS REPORTED TO NOT TOUCH ALL AREAS IN THE CHANNEL. IT IS LIKELY THESE EVENTS WERE CAUSED BY INCORRECT REPROCESSING, WHICH OCCURRED BECAUSE THE USER¿S UNDERSTANDING DIFFERED FROM OLYMPUS RECOMMENDATION. THE USER¿S REPROCESSING STEPS, REPORTED BY ENDOSCOPY SUPPORT SPECIALIST (ESS), DEVIATED FROM REPROCESSING MANUAL AS FOLLOWS: THE USER DOES NOT ALWAYS USE THE AIR/WATER CHANNEL CLEANING ADAPTER DURING PRE-CLEANING. PER THE INSTRUCTIONS FOR USE, ¿TO PREVENT CLOGGING OF THE AIR/WATER NOZZLE OF THE ENDOSCOPE, FLUSH WATER INTO THE AIR CHANNEL OF THE ENDOSCOPE, USING THE AW CHANNEL CLEANING ADAPTER (MH-948) AFTER EACH PATIENT PROCEDURE.¿ THE USER LEAK TESTS THE DEVICE FOR ONLY 10 SECONDS AND DOES NOT ANGULATE THE BENDING SECTION. PER THE INSTRUCTIONS FOR USE, DETAILED STEPS OF LEAK TEST IS MENTIONED AT ¿5.4 LEAKAGE TESTING OF THE ENDOSCOPE¿. ¿WITH THE LEAKAGE TESTER ATTACHED, IMMERSE THE ENDOSCOPE IN THE WATER AND OBSERVE FOR APPROXIMATELY 30 SECONDS WHILE DEFLECTING THE BENDING SECTION OF THE ENDOSCOPE BY TURNING THE ENDOSCOPE¿S UP/DOWN AND RIGHT/LEFT ANGULATION CONTROL KNOBS. CONFIRM THAT THERE IS NO LOCATION ON THE ENDOSCOPE FROM WHICH A CONTINUOUS SERIES OF AIR BUBBLES EMERGES.¿ THE USER DOES NOT REMOVE THE DEVICE FROM WATER PRIOR TO REMOVING THE LEAK TESTER FROM THE DEVICE. THE SCOPE CONNECTOR IS THE ONLY PART REMOVED FROM THE WATER. PER THE INSTRUCTIONS FOR USE, DETAILS OF THE REPROCESSING STEPS ARE MENTIONED AT ¿5.4 LEAKAGE TESTING OF THE ENDOSCOPE¿. ¿REMOVE THE ENDOSCOPE FROM THE WATER WITH THE LEAKAGE TESTER STILL ATTACHED.¿ THE USER DOES NOT TURN OFF MU-1 MAINTENANCE UNIT BEFORE DISCONNECTING LEAK TESTER FROM THE DEVICE. PER THE INSTRUCTIONS FOR USE, DETAILS OF THE LEAK TESTING STEPS AT ¿5.4 LEAKAGE TESTING OF THE ENDOSCOPE¿. ¿DETACH THE LEAKAGE TESTER FROM THE MAINTENANCE UNIT (MU-1) BEFORE DETACHING THE LEAKAGE TESTER FROM THE ENDOSCOPE. IF THE LEAKAGE TESTER IS DETACHED FROM THE ENDOSCOPE BEFORE DETACHING THE LEAKAGE TESTER FROM THE MAINTENANCE UNIT, THE AIR PRESSURE INSIDE THE ENDOSCOPE WILL NOT VENT PROPERLY. THIS MAY DAMAGE THE ENDOSCOPE.¿ THE USER DOES NOT FOLLOW THE CORRECT BRUSHING ORDER DURING MANUAL CLEANING. PER THE INSTRUCTIONS FOR USE, TE BRUSHING STEPS ARE MENTIONED AT ¿5.5 MANUALLY CLEANING THE ENDOSCOPE AND ACCESSORIES BRUSH THE CHANNELS¿. THE USER DOES NOT WIPE THE DEVICE DOWN AFTER FLUSHING WITH DETERGENT SOLUTION AND DO NOT LET IT SOAK FOR THE DETERGENT MANUFACTURER'S RECOMMENDED TIME. PER THE INSTRUCTIONS FOR USE, ¿DRY EXTERNAL SURFACES: DRY THE EXTERNAL SURFACES OF THE ENDOSCOPE, THE CHANNEL PLUG, THE INJECTION TUBE, AND THE AUXILIARY WATER TUBE BY WIPING WITH CLEAN LINT-FREE CLOTHS. 5.5 MANUALLY CLEANING THE ENDOSCOPE AND ACCESSORIES FILL A CLEAN, LARGE BASIN WITH THE DETERGENT SOLUTION AT THE TEMPERATURE AND CONCENTRATION RECOMMENDED BY THE DETERGENT MANUFACTURER.¿ THE DEVICE IS NOT SOAKED AT ALL AND GOES DIRECTLY INTO THE RINSE SINK. PER THE INSTRUCTIONS FOR USE, ¿5.5 MANUALLY CLEANING THE ENDOSCOPE AND ACCESSORIES FILL A CLEAN, LARGE BASIN WITH THE DETERGENT SOLUTION AT THE TEMPERATURE AND CONCENTRATION RECOMMENDED BY THE DETERGENT MANUFACTURER. ¿ OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO CORRECT THE PRODUCT CODE FOR THE SUBJECT DEVICE FROM FDF TO FDS. SEE SECTION D2.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO CORRECT THE LEGAL MANUFACTURER'S CONTACT INFORMATION AND FACILITY REGISTRATION NUMBER. THE FACILITY REGISTRATION NUMBER IS 9610595.

Description of Event or Problem · 0

AS REPORTED FOR THIS EVENT BY THE CUSTOMER, CONCERN WAS RAISED FOR THE REPROCESSING ACROSS MULTIPLE CUSTOMER SITES UPON FINDINGS AFTER DEVICES WERE INSPECTED WITH A BORESCOPE. THE FINDINGS OF A BORESCOPE INSPECTION FOR THIS DEVICE ARE MINOR BUCKLING AT BOOT, WHITE FLECK SEDIMENT AT SUCTION PORT, AND THREADS AT DISTAL TIP. THERE IS NO REPORTED HARM TO ANY PATIENT OR PERSONS. CUSTOMER HAS RE-CLEANED THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2660509 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE FLEXIBLE VIDEO GASTRODUODENOSCOPE FDS AIZU OLYMPUS CO., LTD. GIF-H190 04953170305290
721622 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE FLEXIBLE VIDEO GASTRODUODENOSCOPE FDS AIZU OLYMPUS CO., LTD. GIF-H190 04953170305290

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown