SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2022-02906
- Event Type
- Malfunction
- Date Received
- July 12, 2022
- Date of Event
- June 16, 2022
- Report Date
- July 12, 2022
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K133801
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
THE DEVICE WAS RECEIVED FOR EVALUATION. DURING FUNCTIONAL TESTING, THE REPORTED PROBLEM 'TAKES TOO LONG TO ADMINISTER DRUGS' WAS REPRODUCED. A SERVICE HISTORY REVIEW WAS PERFORMED AND REVEALED THAT THE DEVICE HAS NO PREVIOUS SERVICE EVENTS; THEREFORE, SERVICING DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE A FAILED CAM SHAFT. THE CAM SHAFT REQUIRES REPLACEMENT TO ADDRESS THIS ISSUE. DURING EVALUATION THE DEVICE HAD DELAYED KEYPAD RESPONSE. THE CAUSE WAS IDENTIFIED TO BE A FAILED PROCESSOR BOARD. THE PROCESSOR BOARD REPLACEMENT TO CORRECT THE CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SPECTRUM PUMP TOOK TOO LONG TO ADMINISTER DRUGS AT 954935; PBLM DES. THE STAFF COMPLAINS THAT IT TOOK TOO LONG TO ADMINISTER DRUGS AT 590/HOUR. THE MEDICATION WAS CHEMO AT ROOM TEMPERATURE. VOLUME TO BE INFUSED WAS APPROXIMATELY 270 ML AT 590 ML, DOSE WAS UNKNOWN. NO INFORMATION ON THE IV SET. THE BAGS COME FROM THE PHARMACY READY TO INFUSE. PRIMARY WAS USED. NO ADD-ONS TO THE IV SET. PATIENT IMPACT WAS THEIR TIME. UNDER INFUSION WAS DETERMINED BY THE LENGTH OF TIME IT TOOK TO DELIVER THE MEDICATION. DISCOVERED THE SAME DAY BECAUSE IT WAS A TIMED DOSE AND TOOK TOO LONG. THE EVENT HAPPENED DURING PATIENT INFUSION AT THE ONCOLOGY DEPARTMENT. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1292766 | SPECTRUM INFUSION PUMP | PUMP, INFUSION | FRN | BAXTER HEALTHCARE CORPORATION | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |