MDR report key: 14989106 · Received July 12, 2022

Report

Report Number
2247858-2022-00104
Event Type
Malfunction
Date Received
July 12, 2022
Date of Event
June 14, 2022
Report Date
November 28, 2022
Manufacturer
BOLTON MEDICAL, INC.
Product Code
MIH
PMA / PMN Number
P190015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

BOLTON MEDICAL IS VOLUNTARILY REPORTING AN EVENT RELATED TO A TREO ABDOMINAL STENT-GRAFT CUSTOM-MADE DEVICE. THE CUSTOM MADE TREO DEVICE IS NOT MARKETED IN THE US, HOWEVER IT IS SIMILAR TO THE TREO ABDOMINAL STENT GRAFT SYSTEM APPROVED FOR SALE IN THE US (P190015). THE EVENT OCCURRED IN THE NETHERLANDS.

Additional Manufacturer Narrative · 0

BOLTON MEDICAL IS VOLUNTARILY REPORTING AN EVENT RELATED TO A TREO ABDOMINAL STENT-GRAFT CUSTOM-MADE DEVICE. THE CUSTOM MADE TREO DEVICE IS NOT MARKETED IN THE US, HOWEVER IT IS SIMILAR TO THE TREO ABDOMINAL STENT GRAFT SYSTEM APPROVED FOR SALE IN THE US (P190015). THE EVENT OCCURRED IN THE NETHERLANDS.

Description of Event or Problem · 0

HOLE IN SHEATH WHICH CAUSED STENT PROTRUDING THE SHEATH A BIT WHICH GAVE A LOT OF FRICTION WHEN RELEASING THE STENT. THIS EXTENSIVE AMOUNT OF FRICTION BROKE THE SHEATH RELEASE LEVER AND DEPLOYMENT WAS NOT POSSIBLE. PATIENT OUTCOME: "UNTREATED RIGHT ILIAC COMMON ARTERY ANEURYSM, NO ADDITIONAL INJURIES INFLICTED DUE TO FAILED DEPLOYMENT."

Description of Event or Problem · 0

HOLE IN SHEATH WHICH CAUSED STENT PROTRUDING THE SHEATH A BIT WHICH GAVE A LOT OF FRICTION WHEN RELEASING THE STENT. THIS EXTENSIVE AMOUNT OF FRICTION BROKE THE SHEATH RELEASE LEVER AND DEPLOYMENT WAS NOT POSSIBLE. PATIENT OUTCOME - "UNTREATED RIGHT ILIAC COMMON ARTERY ANEURYSM, NO ADDITIONAL INJURIES INFLICTED DUE TO FAILED DEPLOYMENT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
652729 TREO ABDOMINAL STENT-GRAFT SYSTEM STENT, ENDOVASCULAR GRAFT, AORTIC MIH BOLTON MEDICAL, INC. 2205260076

Patients

Seq Age Sex Outcome Treatment
1 74 YR Male