STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2022-11760
- Event Type
- Injury
- Date Received
- July 12, 2022
- Date of Event
- March 1, 2022
- Report Date
- September 6, 2022
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- UDI-DI
- 10888628000223
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5, D6B, D9, H3, H6.
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON AUGUST 08, 2022, WITH LOT NUMBER 2925416. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: OPENING ON POSTERIOR, WEAR ABRASION AND CREASE FOLD. LEAK TEST AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED: CREASE SMOOTH OPENING ON POSTERIOR, YELLOW BIOLOGICAL TISSUE INNER THE DEVICE AND OBSERVED VOID >1 MM IN NON-TEXTURED, VALVE-TO SHELL-BOND AREA. THE FILL TEST INSPECTION WAS PERFORMED, THE RESULT IS NO BLOCKAGE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: OPENING CREASE SMOOTH ON POSTERIOR ASSESSED AS FOLD FLAW OPENING.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.
DEVICE HAS BEEN EXPLANTED
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 701564 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FWM | ALLERGAN (COSTA RICA) | 68-330 | 2925416 | 10888628000223 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR | Female | Required Intervention |