FDA Adverse Event Malfunction Summary report: N

640G INSULIN PUMP MMT-1712W

MDR report key: 14975533 · Received July 11, 2022

Report

Report Number
2032227-2022-280256
Event Type
Malfunction
Date Received
July 11, 2022
Date of Event
September 1, 2017
Report Date
July 11, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
Removal / Correction Number
Z-0958-2020
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. CORRECTION/REMOVAL REPORTING NUMBER: Z-0958-2020. (B)(4). S/W VER 2.9D PUMP NOT IN MANUFACTURE MODE. PUMP PASSED THE SELF TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, AND THE DISPLACEMENT TEST. A WATER FILLED RESERVOIR WAS USED DURING TESTING. SEVERAL BOLUS DELIVERIES WERE PROGRAMMED AND DELIVERED. OBSERVED LIQUID EXIT THE TUBING DURING THE BOLUS DELIVERIES. ALL BOLUSES DELIVERED PROPERLY AND WERE LISTED IN THE DAILY HISTORY SCREEN. NO DELIVERY ANOMALIES NOTED DURING TESTING. PUMP RECEIVED WITH SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, FADED SERIAL NUMBER LABEL, CRACKED RETAINER, AND MINOR SCRATCHED LCD WINDOW. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ""DEFECTS"" OR HAS ""MALFUNCTIONED"". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT."

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER EXPERIENCED HIGH BLOOD GLUCOSE OF 356 MG/DL. CUSTOMER EXPERIENCED NO SYMPTOMS. CUSTOMER TREATED THEMSELVES WITH SYRINGE. BOLUS MATCHED WITH DAILY TOTALS. INSULIN EXITED TUBING. HIGH PRESSION TEST WAS PERFORMED AND GOT 2.1 UNIT RESULT. INSULIN PUMP WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2016010 640G INSULIN PUMP MMT-1712W PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1712W

Patients

Seq Age Sex Outcome Treatment
1 Unknown