ELECTROSURGICAL GENERATOR "ESG-400"
Report
- Report Number
- 9610773-2022-00261
- Event Type
- Malfunction
- Date Received
- July 11, 2022
- Date of Event
- June 29, 2022
- Report Date
- July 11, 2022
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
THE SUSPECT MEDICAL DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR EVALUATION/INVESTIGATION BUT TO AN OLYMPUS REGIONAL REPAIR CENTER (RRC) IN CHINA. THE EVALUATION/INVESTIGATION CONFIRMED THE OCCURRENCE OF ERROR E433. THIS ERROR MESSAGE, WHICH IN THE CASE AT HAND WAS CAUSED BY A DEFECT OF THE HIGH VOLTAGE POWER SUPPLY (HVPS) BOARD, IS TRIGGERED BY THE GENERATOR¿S SAFETY SYSTEM. IN CASE OF CRITICAL ERRORS, THE SAFETY SYSTEM WILL NOT PERMIT ANY FURTHER USE OF THE GENERATOR UNTIL THE ERROR IS RECTIFIED. THUS, THIS EVENT/INCIDENT WAS ATTRIBUTED TO COMPONENT FAILURE. A MANUFACTURING AND QUALITY CONTROL REVIEW WAS PERFORMED FOR THE AFFECTED SERIAL NUMBER OF THE HF GENERATOR WITHOUT SHOWING ANY ABNORMALITIES. THE CASE WILL BE CLOSED ON OLYMPUS SIDE WITH NO FURTHER ACTIONS, BUT THE REPORTED EVENT/INCIDENT WILL BE RECORDED FOR TRENDING AND SURVEILLANCE PURPOSES. FURTHERMORE, THE USER WILL BE INFORMED ABOUT THE INVESTIGATION RESULTS. PLEASE NOTE: THIS REPORT IS BEING SUBMITTED ALTHOUGH THE SUSPECT MEDICAL DEVICE IS NOT MARKETED IN THE USA. HOWEVER, A SIMILAR DEVICE IS MARKETED. MODEL # / CATALOG #: WB91051W; BRAND NAME: ELECTROSURGICAL GENERATOR "ESG-400"; COMMON DEVICE NAME: HF-GENERATORS; 510(K): K203682; PRODUCT CODE: GEI.
OLYMPUS WAS INFORMED THAT AN HF ELECTROSURGICAL GENERATOR "ESG-400" DISPLAYED ERROR MESSAGE E433. SINCE THE AFFECTED DEVICE WAS A LOANER DEVICE AND THE OCCURRENCE OF THE ERROR MESSAGE WAS DETECTED INTERNALLY , NO PROCEDURE WAS AFFECTED AND CONSEQUENTLY NO PATIENT WAS INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1499308 | ELECTROSURGICAL GENERATOR "ESG-400" | HF-GENERATORS | GEI | OLYMPUS WINTER & IBE GMBH | WB91051C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |