CHECK-IT
Report
- Report Number
- 3016521623-2022-00251
- Event Type
- Malfunction
- Date Received
- July 8, 2022
- Date of Event
- June 24, 2022
- Report Date
- July 6, 2022
- Manufacturer
- LUCIRA HEALTH, INC.
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 501
Narratives
PRODUCT HAS NOT BEEN RETURNED. THEREFORE, A DHR REVIEW WAS COMPLETED FOR KIT LOT NUMBER K08A111104224M1 WITH 0 DISCREPANCIES FOUND (ALL LOTS PASSED IN TOTAL PER THE SAMPLING PLAN). A REVIEW OF EXISTING CAPAS, SCARS AND NCMRS WERE COMPLETED AND THERE ARE NO PRIOR RECORDS RELATED TO "FALSE POSITIVE" FAILURE MODE FOR THIS LOT. THERE ARE 0 ADDITIONAL COMPLAINTS FROM THIS CUSTOMER ASSOCIATED WITH A "FALSE POSITIVE" FAILURE MODE, PRIOR TO THE REPORTED RECEIPT DATE OF JUNE 24, 2022. BASED ON REVIEW OF THE PRODUCT'S FMEAS AND RISK ASSESSMENT DOCUMENTATION, FALSE POSITIVE TEST RESULTS ARE A KNOWN POSSIBLE OUTCOME REGARDING THIS ISSUE UNDER EVALUATION, REFER TO FMEA-001 AND FMEA-004. (B)(4). LUCIRA HEALTH WILL CONTINUE TO MONITOR TRENDS RELATED TO FALSE POSITIVE RESULTS IN ACCORDANCE WITH POST-MARKET SURVEILLANCE PROCESS. DHR REVIEW KIT LOT# K08A111104224M1: SAMPLE VIAL LOT DHR REVIEW: 2203438, 2202318, 2203506, 2203231, 2203332 (ASSOCIATED INTERNAL LOT # 220305-1J, 220124-1N, 220305-4D, 220124-7L, 220124-7R) TEST LOT DHRS REVIEW: 2203141, 2203143, 2203141, 2203269, 2203280, 2203282, 2203282, 2203381, 2203448, 2203478, 2203503, 2203503, 2203478, 2203502, 2203502, 2203502, 2203526, 2203577, 2203527, 2203527 (ASSOCIATED INTERNAL LOT # 220305-6N, 220305-6S, 220305-6M, 220305-12N, 220305-12U, 220305-12Y, 220305-12Z, 220305-18F, 220305-21J, 220305-24E, 220305-24L, 220305-24N, 220305-24G, 220305-24I, 220305-24K, 220305-24J, 220305-24X, 220305-27P, 220305-3H, 220305-3I) BASED ON THE LIMITED INFORMATION AVAILABLE, ROOT CAUSE CANNOT BE DETERMINED. POTENTIAL ROOT CAUSES INCLUDE BUT ARE NOT LIMITED TO: ENVIRONMENTAL CONTAMINATION, LOW VIRAL LOAD, AND/OR DEVICE FAILURE. A SUPPLEMENTAL REPORT WILL BE FILED IF ANY FURTHER INVESTIGATION AND/OR ADDITIONAL INFORMATION IS OBTAINED. BASED ON THE COMPLAINT REVIEW, NO HARM WAS REPORTED WITHIN THE COMPLAINT. THIS DEVICE IS MARKETED UNDER EUA: (B)(4) CHECK-IT.
ONE DEVICE REPORTED AS HAVING AN ALLEGED FALSE POSITIVE RESULT. THE COMPLAINANT TOOK AN ADDITIONAL LUCIRA TEST AND ACCULA (RT-PCR) TEST WITH NEGATIVE RESULTS WITHIN 30 MINUTES FOR CONFIRMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2865269 | CHECK-IT | COVID-19 TEST KIT | QJR | LUCIRA HEALTH, INC. | K08A111104224M1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |