FDA Adverse Event Malfunction Summary report: N

BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER

MDR report key: 14960343 · Received July 8, 2022

Report

Report Number
9610048-2022-00068
Event Type
Malfunction
Date Received
July 8, 2022
Date of Event
June 9, 2022
Report Date
September 28, 2022
Manufacturer
BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
Product Code
FOZ
UDI-DI
00382903818235
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL REPORTER PHONE #: (B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 09AUG2022. BD RECEIVED A USED CATHETER FROM AN UNKNOWN LOT FOR EVALUATION. THE CATHETER WAS RECEIVED IN AN OPEN PACKAGE AND NOT THE ORIGINAL PRODUCT PACKAGING. A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED FOR THE REPORTED LOT, 1111254, AND NO QUALITY ISSUES WERE FOUND DURING PRODUCTION. OUR QUALITY ENGINEER VISUALLY INSPECTED THE RETURNED CATHETER AND IDENTIFIED THAT THE CATHETER WAS NOT A BD PRODUCT. ALSO, TWO PHOTOS OF THE DEFECT WERE PROVIDED BUT NOT ENOUGH EVIDENCE WAS VISIBLE IN THE PROVIDED PHOTOS TO DETERMINE ANY POSSIBLE DEFECTS. THEREFORE, BASED OFF THE VISUAL INSPECTION THE ENGINEER COULD NOT VERIFY THIS AS A MANUFACTURING DEFECT. SINCE, THE DEFECT WAS DETERMINED TO NOT BE A BD PRODUCT A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. THE MANUFACTURING FACILITY HAS BEEN NOTIFIED OF THIS INCIDENT AND THE FINDINGS. WERE RECEIVED AT BD JUIZ DE FORA ON 09-AUG2022, 01 SAMPLE USED, CONTAINING ONLY THE CATHETER, IN OPEN PACKAGING AND NOT ORIGINAL PRODUCT. THE SAMPLE WAS RECEIVED BY THE PERSON RESPONSIBLE FOR THE ANALYSIS AND SENT FOR DECONTAMINATION. ACCORDING TO THE VISUAL ANALYSIS OF THE SAMPLES, IT CAN BE SEEN THAT THE CATHETER SENT BY THE CUSTOMER IS NOT A BD PRODUCT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER HAD A DEFECT IN THE FILTER, CAUSING FLUID LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE: "ONE UNIT OF THE BATCH HAD A DEFECT IN THE FILTER, CAUSING FLUID LEAKAGE.¿

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER HAD A DEFECT IN THE FILTER, CAUSING FLUID LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE: "ONE UNIT OF THE BATCH HAD A DEFECT IN THE FILTER, CAUSING FLUID LEAKAGE.¿.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2866375 BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER INTRAVASCULAR CATHETER FOZ BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. 381823 1111254 00382903818235

Patients

Seq Age Sex Outcome Treatment
1 Unknown