BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE
Report
- Report Number
- 1917413-2022-00377
- Event Type
- Malfunction
- Date Received
- July 8, 2022
- Date of Event
- June 10, 2022
- Report Date
- June 20, 2022
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- JKA
- UDI-DI
- 50382903627610
- PMA / PMN Number
- K911731
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: MATERIAL #: 362761. LOT/BATCH #: 1232566. BD HAD NOT RECEIVED SAMPLES, BUT 1 PHOTO WAS PROVIDED FOR INVESTIGATION. THE PHOTO WAS REVIEWED AND THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS OBSERVED. ADDITIONALLY, RETENTION SAMPLES OF THE INCIDENT LOT WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS NOT OBSERVED. NO DIFFICULTIES WERE ENCOUNTERED DURING BLOOD COLLECTION AND ALL TUBES APPEARED TO EXHIBIT PROPER FILL. BD WAS UNABLE TO DUPLICATE THE CUSTOMER¿S INDICATED FAILURE (POOR BARRIER) BECAUSE THE DEFECT WAS NOT EVIDENT IN THE TESTING OF THE COMPLAINT LOT SAMPLES. REPLICATES OF BOTH RETAIN AND CONTROL SAMPLES TESTED DEMONSTRATED CLINICALLY ACCEPTABLE PERFORMANCE FOR ALL VISUAL OBSERVATIONS EVALUATED. ALL TUBES PERFORMED AS EXPECTED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE POOR BARRIER SEPARATION BASED ON PHOTOS ONLY. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED WHEN USING THE BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE THE SAMPLE WAS NOT SEPARATED PROPERLY. THIS EVENT OCCURRED 2 TIMES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: ¿THIS IS A REPORT ABOUT POOR SEPARATION OF SAMPLE. ACCORDING TO THE CUSTOMER'S REPORT, SAMPLE OPERATION WAS PERFORMED IN ACCORDANCE WITH THE PROTOCOL BUT THE SAMPLE WAS NOT SEPARATED PROPERLY."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1986583 | BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO. (BROKEN BOW) | 362761 | 1232566 | 50382903627610 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |