IMP,TM 3.7MM MTX,10MM
Report
- Report Number
- 0002023141-2022-01755
- Event Type
- Injury
- Date Received
- July 8, 2022
- Date of Event
- June 10, 2022
- Report Date
- December 15, 2022
- Manufacturer
- ZIMMER DENTAL
- Product Code
- DZE
- UDI-DI
- 00889024018860
- PMA / PMN Number
- K132258
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMMER BIOMET COMPLAINT NUMBER (B)(4). PRODUCT HAS BEEN RECEIVED BY ZIMMER BIOMET AND THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
ZIMMER BIOMET COMPLAINT NUMBER (B)(4) THE FOLLOWING SECTIONS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT B5: DESCRIBE EVENT OR PROBLEM G3: DATE RECEIVED BY THE MANUFACTURER H1: TYPE OF REPORT, FOLLOW-UP NUMBER H2: FOLLOW UP TYPE H3: DEVICE EVALUATED BY MANUFACTURER: CHANGE ¿NO' TO 'YES' H6: EVALUATION CODES H10: ADDITIONAL NARRATIVE ONE IMPLANT WAS RETURNED FOR INVESTIGATION. VISUAL EVALUATION OF THE AS RETURNED IMPLANT IDENTIFIED IMPLANT FRACTURED AT THE COLLAR REGION. SIGNS OF USE. FUNCTIONAL TESTING COULD NOT BE PERFORMED SINCE THE PRODUCT WAS FRACTURED. PRE-EXISTING CONDITIONS NOTED ON THE PER WERE 'LOW BONE DENSITY ¿ TYPE III'. THE REPORTED DEVICE HAD BEEN PLACED ON TOOTH 32 (FDI) FOR APPROXIMATELY 5 MONTHS. THE PATIENT IS A SMOKER X-RAY & PICTURE EVALUATION: PICTURE OR X-RAY EVIDENCE NOT PROVIDED. REVIEW OF APPROPRIATE DOCUMENTATION: DOCUMENTS REVIEWED: INSTRUCTIONS FOR USE - TAPERED SCREW-VENT AND TRABECULAR METAL IMPLANTS, IFU4869 REV 9-10/19. INFORMATION IDENTIFIED: CONTRAINDICATIONS & PRECAUTIONS. PER THE APPLICABLE IFU, IT IS STATED THAT IMPROPER TECHNIQUES, SEVERE BRUXISM, CLENCHING, AND OVERLOADING, MAY CAUSE COMPONENT FRACTURE. DHR REVIEW: DHR REVIEW FOR THE LOT (1232699) HAD REVEALED NO DEVIATIONS NOR NON-CONFORMANCES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENTS. ALL PRODUCTS WERE CONFORMING AT THE TIME THEY LEFT ZIMVIE. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW: COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (1232699) FOR SIMILAR EVENTS (COMPLAINT CATEGORY KEYWORDS: FRACTURE: IMPLANT) AND NO OTHER COMPLAINT WAS IDENTIFIED. POST MARKET TRENDING REVIEW: OCTOBER POST MARKET TRENDING WAS REVIEWED AND THERE WERE NO ACTIONABLE EVENTS OR CORRECTIVE ACTIONS FOR THE REPORTED EVENT OR PRODUCT. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR, AND THE REPORTED EVENT WAS CONFIRMED.
DOCTOR REPORTED THAT AN IMPLANT FRACTURE AT TOOTH SITE #32.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2851146 | IMP,TM 3.7MM MTX,10MM | DENTAL IMPLANT | DZE | ZIMMER DENTAL | TMMB10 | 1232699 | 00889024018860 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Female | Required Intervention |