FDA Adverse Event Malfunction Summary report: N

INFINITY RETRO-REAMER, 10MM

MDR report key: 14955694 · Received July 8, 2022

Report

Report Number
1017294-2022-00077
Event Type
Malfunction
Date Received
July 8, 2022
Date of Event
June 21, 2022
Report Date
August 4, 2022
Manufacturer
CONMED LARGO
Product Code
GFG
UDI-DI
20845854902025
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED DEVICE IS BEING RETURNED TO CONMED FOR EVALUATION. A SUPPLEMENTAL AND FINAL REPORT WILL BE FILED FOLLOWING THE COMPLETION OF THE DEVICE EVALUATION AND COMPLAINT INVESTIGATION. THIS ISSUE WILL CONTINUE TO BE MONITORED THROUGH THE COMPLAINT SYSTEM TO ASSURE PATIENT SAFETY.

Additional Manufacturer Narrative · 0

CORRECTION: D4.LOT NUMBER WAS CORRECTED FROM 1232757 TO LOT NUMBER:1230761 AS NEW INFORMATION WAS PROVIDED SINCE INITIAL FILING. MANUFACTURER NARRATIVE: CUSTOMER EVENT ¿REAMER TIP SHATTERED INSIDE PATIENT¿ WAS CONFIRMED BASED ON PHOTOGRAPHIC EVIDENCE AND DEVICE EVALUATION. VISUAL EXAMINATION OF THE RETURNED DEVICE ITEM KRR100, FOUND DEVICE BROKEN OFF AT THE TIP, BROKEN PIECE WAS NOT RETURNED FOR EVALUATION. ALSO, EVALUATION FOUND REAMER CAP BROKEN OFF. THE MANUFACTURING DOCUMENTS FROM THE DEVICE HISTORY RECORD HAVE BEEN REVIEWED WITH SPECIAL ATTENTION TO THE MANUFACTURING AND INSPECTION OF THE PRODUCT. THE PRODUCT RELEASED FOR DISTRIBUTION WAS FOUND TO HAVE MET ALL SPECIFICATIONS PRIOR TO SHIPMENT. A TWO-YEAR LOT HISTORY REVIEW WAS CONDUCTED AND FOUND A TOTAL OF 2 SIMILAR EVENTS INVOLVING 2 DEVICES FOR THIS LOT NUMBER. A TWO-YEAR REVIEW OF COMPLAINT HISTORY REVEALED THERE HAS BEEN A TOTAL OF 5 COMPLAINTS, REGARDING 5 DEVICES, FOR THIS DEVICE FAMILY AND FAILURE MODE. DURING THIS SAME TIME FRAME (B)(4). PER THE INSTRUCTIONS FOR USE, THE USER IS ADVISED DO NOT USE EXCESSIVE FORCE ON INSTRUMENTS TO AVOID DAMAGE OR BREAKAGE DURING USE. DO NOT USE INSTRUMENTS TO PRY, AS BENDING OR BREAKAGE MAY OCCUR. EXERCISE CARE IN THE USE OF THE HANDPIECE HOLDING THE REAMER TO MINIMIZE SIDE OR BENDING LOADS TO THE REAMER WHICH MAY CAUSE REAMER BREAKAGE AND/OR OVERSIZED TUNNELS. EXCESSIVE FORCE APPLIED DURING RETROGRADE DRILLING CAN LEAD TO REDUCED SOCKET SIZES. THIS ISSUE WILL CONTINUE TO BE MONITORED THROUGH THE COMPLAINT SYSTEM TO ASSURE PATIENT SAFETY.

Description of Event or Problem · 0

THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE KRR100, INFINITY RETRO-REAMER, 10MM WAS BEING USED DURING AN ACL RECONSTRUCTION ON (B)(6) 2022 WHEN IT WAS REPORTED THAT THE ¿REAMER TIP SHATTERED INSIDE PATIENT WHEN RETRO REAMING THE FEMORAL SOCKET. THE DOCTOR HAD TO FISH OUT 2 PIECES OF THE TIP FROM INSIDE THE JOINT.¿ THE PROCEDURE WAS COMPLETED WITH REPORT OF A 10-MINUTE DELAY TO THE CASE. FURTHER ASSESSMENT FOUND THE REAMER TIP WAS REMOVED FROM INSIDE THE JOINT USING A GRASPER. THERE WAS NO INJURY TO THE PATIENT AND NO MEDICAL INTERVENTION OR EXTENDED HOSPITALIZATION WAS REQUIRED FOR THE PATIENT. THE PATIENT IS SAID TO BE OKAY AND UNHARMED. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

Description of Event or Problem · 0

THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE KRR100, INFINITY RETRO-REAMER, 10MM WAS BEING USED DURING AN ACL RECONSTRUCTION ON (B)(6) 22 WHEN IT WAS REPORTED THAT THE ¿REAMER TIP SHATTERED INSIDE PATIENT WHEN RETRO REAMING THE FEMORAL SOCKET. THE DOCTOR HAD TO FISH OUT 2 PIECES OF THE TIP FROM INSIDE THE JOINT.¿ THE PROCEDURE WAS COMPLETED WITH REPORT OF A 10-MINUTE DELAY TO THE CASE. FURTHER ASSESSMENT FOUND THE REAMER TIP WAS REMOVED FROM INSIDE THE JOINT USING A GRASPER. THERE WAS NO INJURY TO THE PATIENT AND NO MEDICAL INTERVENTION OR EXTENDED HOSPITALIZATION WAS REQUIRED FOR THE PATIENT. THE PATIENT IS SAID TO BE OKAY AND UNHARMED. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1986297 INFINITY RETRO-REAMER, 10MM BIT, SURGICAL GFG CONMED LARGO KRR100 1230761 20845854902025

Patients

Seq Age Sex Outcome Treatment
1 Unknown