FDA Adverse Event Injury Summary report: N

TSOLUTION ONE TOTAL KNEE APPLICATION

MDR report key: 14948022 · Received July 7, 2022

Report

Report Number
3000719653-2019-00001
Event Type
Injury
Date Received
July 7, 2022
Date of Event
July 12, 2019
Report Date
July 7, 2022
Manufacturer
THINK SURGICAL INC.
Product Code
OLO
PMA / PMN Number
K191369
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DURING REMOVAL OF FIXATION PINS AFTER SURGERY WAS COMPLETED, THE FEMUR FIXATION PIN SNAPPED OFF BY THE THREADS. THE BROKEN END OF PIN WAS SUCCESSFULLY REMOVED FROM THE PATIENT.UNDER MAGNIFICATION THERE DOES NOT APPEAR TO BE ANY MATERIAL DEFECTS AT THE LOCATION OF THE BREAK, SUCH AS AIR POCKETS OR HAIRLINE CRACKS AND THE PIN DOES NOT APPEAR TO BE BENT OR DEFORMED. THE PIN WAS REMOVED WITH A BROKEN SCREW REMOVAL KIT AND NEEDLE NOSE PLIERS. THE PIN WAS REMOVED SUCCESSFULLY WITHOUT ADDITIONAL INTERVENTION. NO FURTHER CORRECTIVE ACTION IS NECESSARY. WE WILL MONITOR VIA THE COMPLAINT PROCESS FOR ADDITIONAL INSTANCES OF BROKEN PINS AND WILL TAKE ACTION AS WARRANTED.

Description of Event or Problem · 0

DURING REMOVAL OF FIXATION PINS AFTER SURGERY WAS COMPLETED, THE FEMUR FIXATION PIN WAS BEING REMOVED BY HAND WITH OUR T-HANDLE WRENCH 101884 AND SNAPPED OFF BY THE THREADS. THE BROKEN END OF PIN WAS SUCCESSFULLY REMOVED FROM THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1598514 TSOLUTION ONE TOTAL KNEE APPLICATION TCAT OLO THINK SURGICAL INC. 103306 12091

Patients

Seq Age Sex Outcome Treatment
1 Unknown