BD 10ML SYRINGE LUER-LOK¿ TIP
Report
- Report Number
- 1213809-2022-00368
- Event Type
- Malfunction
- Date Received
- July 6, 2022
- Date of Event
- June 6, 2022
- Report Date
- July 14, 2022
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- UDI-DI
- 30382903096450
- PMA / PMN Number
- K110771
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVAL YES, D9: RETURNED TO MANUFACTURER ON: 27-JUN-2022. H6: INVESTIGATION SUMMARY: ONE SAMPLE AND TWO PHOTOS OF A 10ML LUER-LOCK SYRINGE (P/N 309645) WERE RECEIVED AND EVALUATED. THE SAMPLE HAS A CRACK IN THE BARREL MEASURING 11/8 INCH LONG STARTING ¼ INCH FROM THE ROOF OF THE BARREL. THE CONDITION OBSERVED IS NON-CONFORMING PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE CRACKED BARREL DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1113346. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.
IT WAS REPORTED THAT WHILE USING BD 10ML SYRINGE LUER-LOK¿ TIP PRODUCT LEAKED OUT A SPLIT/CRACK ON THE SIDE OF TWO SYRINGES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DATE: (B)(6) 2022 COMPLAINT NUMBER: (B)(4). ITEM NUMBER: 308-309645 ITEM DESCRIPTION: 10ML LL STER SYRINGE W/ 20GX1.5¿? NDL REG INCIDENT DESCRIPTION: AS PER CS, THIS CUSTOMER ADVISED THAT THE BELOW SYRINGES APPEAR TO HAVE A SPLIT/CRACK IN THE SIDES. WHEN PUSHING MEDICATION THROUGH THE SYRINGE PRODUCT LEAKS OUT THE SIDE. THIS HAS HAPPEN WITH 2 DIFFERENT SYRINGES.
IT WAS REPORTED THAT WHILE USING BD 10ML SYRINGE LUER-LOK¿ TIP PRODUCT LEAKED OUT A SPLIT/CRACK ON THE SIDE OF TWO SYRINGES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DATE: 06/06/2022, COMPLAINT NUMBER: (B)(4), ITEM NUMBER: 308-309645, ITEM DESCRIPTION: 10ML LL STER SYRINGE W/ 20GX1.5¿? NDL REG. INCIDENT DESCRIPTION: AS PER CS, THIS CUSTOMER ADVISED THAT THE BELOW SYRINGES APPEAR TO HAVE A SPLIT/CRACK IN THE SIDES. WHEN PUSHING MEDICATION THROUGH THE SYRINGE PRODUCT LEAKS OUT THE SIDE. THIS HAS HAPPEN WITH 2 DIFFERENT SYRINGES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 966408 | BD 10ML SYRINGE LUER-LOK¿ TIP | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 309645 | 1113346 | 30382903096450 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |