FDA Adverse Event Malfunction Summary report: N

RELION® INSULIN SYRINGE

MDR report key: 14940324 · Received July 6, 2022

Report

Report Number
1920898-2022-00433
Event Type
Malfunction
Date Received
July 6, 2022
Date of Event
June 7, 2022
Report Date
July 28, 2022
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
00681131138376
PMA / PMN Number
K190054
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: CUSTOMER RETURNED A SINGLE 0.5ML SYRINGE. VISUAL INSPECTION OF THE SYRINGE FOUND A CRACK IN THE SYRINGE¿S SURFACE RUNNING DOWN THE BARREL OF THE SYRINGE FROM ROUGHLY THE (B)(4) TO (B)(4) UNIT MARKERS. ADDITIONALLY, WATER WAS FOUND TO LEAK THROUGH THIS CRACK WHEN ATTEMPTING TO DRAW FLUID INTO THE BARREL OF THE SYRINGE. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 1242742. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BASED ON THE SAMPLES RECEIVED, EMBECTA WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF LEAKAGE. A ROOT-CAUSE COULD NOT BE DETERMINED. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE RELION® INSULIN SYRINGE WAS CRACKED AND LEAKED INSULIN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "FOUND 1 SYRINGE WITH A CRACK IN THE BARREL OF SYRINGE WHILE DRAWING UP INSULIN. FOUND INSULIN TO WET HER HAND."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE RELION® INSULIN SYRINGE WAS CRACKED AND LEAKED INSULIN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "FOUND 1 SYRINGE WITH A CRACK IN THE BARREL OF SYRINGE WHILE DRAWING UP INSULIN. FOUND INSULIN TO WET HER HAND."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2722710 RELION® INSULIN SYRINGE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 1242742 00681131138376

Patients

Seq Age Sex Outcome Treatment
1 Unknown