FDA Adverse Event Malfunction Summary report: N

MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 14913425 · Received July 5, 2022

Report

Report Number
2032227-2022-269796
Event Type
Malfunction
Date Received
July 5, 2022
Date of Event
June 11, 2022
Report Date
September 30, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000521554
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

PUMP RECEIVED WITH BLANK DISPLAY. UNABLE TO VERIFY FROZEN DISPLAY OR PERFORM THE DISPLACEMENT TEST, SLEEP CURRENT TEST, ACTIVE CURRENT TEST AND SELF TEST DUE TO BLANK DISPLAY. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE ON THE PCBA 1 / PCBA 2 / FORCE SENSOR / MOTOR / HARNESS ASSEMBLY VIBRATOR / CORRODED BATTERY TUBE. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CORRODED BATTERY TUBE, CRACKED BATTERY TUBE THREADS, CRACKED CASE ALONG THE SIDE OF THE BATTERY TUBE AND MINOR SCRATCHED LCD WINDOW. BLANK DISPLAY DUE TO MOISTURE DAMAGE ON THE PCBA 1 / PCBA 2 / HARNESS ASSEMBLY / CORRODED BATTERY TUBE. UNABLE TO CONFIRMED FROZEN SCREEN DUE TO BLANK DISPLAY. MOISTURE DAMAGE FOUND ON THE PCBA 1 / PCBA 2 / FORCE SENSOR / MOTOR / HARNESS ASSEMBLY VIBRATOR / CORRODED BATTERY TUBE.¿ COSMETIC DAMAGE FOUND ON THE PUMP CASE. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED A BLANK DISPLAY & FROZEN DISPLAY. THE CUSTOMER WAS ON THE BEACH. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. TROUBLESHOOTING WAS PERFORMED. THE PRODUCT WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2148038 MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MG/DL ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1886 HG5DQE5ZZ 000000763000521554

Patients

Seq Age Sex Outcome Treatment
1 Unknown