FDA Adverse Event Malfunction Summary report: N

MCKESSON MEDI-LAB HCG COMBO

MDR report key: 1491073 · Received August 4, 2009

Report

Report Number
2027969-2009-00622
Event Type
Malfunction
Date Received
August 4, 2009
Date of Event
July 13, 2009
Report Date
August 3, 2009
Manufacturer
BIOSITE INCORPORATED
Product Code
JHI
PMA / PMN Number
993065
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION: LOT#(S): HCG8080032. EXP DATE(S): 07/2010. CONTROL LOT: 25MIU/ML HCG URINE, LOT: HCG090107-03. 100MIU/ML HCG URINE, LOT: HCG090521-01. 255.3IU/ML HCG URINE, LOT: HCG 090526-03. SUMMARY OF RESULTS: THE RETENTION DEVICES MEET QC SPECIFICATION. CORRECT POSITIVE RESULTS WERE OBSERVED WHEN TESTED WITH 25MIU/ML HCG URINE CONTROL AT 3 MINUTES READ TIME. THE 100MIU/ML, 255.3IU/ML HCG URINE CONTROL YIELDED CLEARLY POSITIVE RESULTS AT 3 MINUTES READING TIME. THE RETENTION DEVICES MEET QC SPECIFICATIONS. NO FALSE NEGATIVE RESULT WAS OBSERVED; THIS COMPLAINT IS NOT CONFIRMED. RETURN 8 UNUSED DEVICES LOT: HCG8080032 EXP 2010/0. RETURN 1 VIAL RETURN URINE SPECIMEN (0.5 ML). RETENTION DEVICE, LOT: HCG8080032 EXP 2010/07. WHEN RUNNING THE 1 VIAL RETURN URINE SPECIMEN FROM CUSTOMER WITH RETURN AND RETENTION DEVICE (N=1/EA), ALL THE SPECIMEN RESULTS WERE NEGATIVE IN URINE AT THE 3 - MINUTE READ TIME. THERE WAS INSUFFICIENT SAMPLE TO PERFORM HCG QUANT. RETURN URINE SPECIFIC GRAVITY WAS 1.022 AND PH WAS 5.0. FROM THE COMPLAINT INFO, IT IS STATED THAT HCG QUANT IN SERUM SAMPLE WAS 11.5, SO IT IS POSSIBLE TO GET A NEGATIVE RESULT IF THE URINE HCG LEVEL IS LOWER THAN THE CUTOFF (25MIU/ML). NOTHING ABNORMAL FOUND IN BATCH REVIEW AND THE PRODUCT #, PRODUCT DESCRIPTION AND LOT # WAS VERIFIED. NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CUSTOMER REPORTED TWO FALSE NEGATIVE URINE HCG RESULTS ON TWO DIFFERENT PTS. IN 2009, A FEMALE HAD FALSE NEGATIVE RESULTS AND A SERUM HCG QUANT RESULT OF 11.5. SPECIFIC GRAVITY ON THE URINE WAS 1.030.

Description of Event or Problem · 2

CUSTOMER REPORTED TWO FALSE NEGATIVE URINE HCG RESULTS ON TWO DIFFERENT PTS. APPROXIMATELY 6 WKS AGO, A DIFFERENT PT WAS NEGATIVE AT 3' WITH A FAINT LINE AT >3'. URINE WAS SENT TO HOSP LAB WHO GOT NEGATIVE AT 3' AS WELL; SERUM QUANT WAS 30; SPECIFIC GRAVITY WAS LOW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MCKESSON MEDI-LAB HCG COMBO HCG PREGNANCY TEST JHI BIOSITE INCORPORATED FHC-202-KMK25 HCG8080032

Patients

Seq Age Sex Outcome Treatment
1 25 YR
2 NI