INRATIO
Report
- Report Number
- 2027969-2009-00617
- Event Type
- Malfunction
- Date Received
- July 31, 2009
- Date of Event
- July 7, 2009
- Report Date
- July 30, 2009
- Manufacturer
- BIOSITE INCORPORATED
- Product Code
- JPA
- PMA / PMN Number
- 021923
- Removal / Correction Number
- NI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- UNKNOWN
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED. DATE: (B)(6) 2009, INRATIO: 4.1, LAB: 3.3, MEAN: 3.70, CONFIDENCE LIMITS: 2.2-5.3. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. FUNCTIONAL TESTING IS NOT REQUIRED AT THIS TIME; METER AND STRIPS ARE NOT EXPECTED TO BE RETURNED. THIS FAILURE MODE WILL CONTINUE TO BE MONITORED TO DETERMINE IF FUTURE CORRECTIVE ACTION IS WARRANTED. AS OF 07/15/2009, 25 DISCREPANT RESULTS COMPLAINTS WERE REPORTED FOR LOT #080868 YIELDING A COMPLAINT RATE OF 0.008%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (>0.07%) NO FURTHER ACTION IS REQUIRED AT THIS TIME. ON GOING TRENDING SHALL BE PERFORMED TO DETERMINE IF FURTHER ACTION IS WARRANTED. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS:" DATE: (B)(6) 2009, INRATIO: 4.1, LAB: 3.3.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | BIOSITE INCORPORATED | 0100007 | 080868 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |