FDA Adverse Event Malfunction Summary report: N

SPECTRUM IQ INFUSION PUMP

MDR report key: 14880715 · Received June 30, 2022

Report

Report Number
1314492-2022-02754
Event Type
Malfunction
Date Received
June 30, 2022
Report Date
July 26, 2022
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
UDI-DI
00085412610900
PMA / PMN Number
K220417
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IT WAS REPORTED THAT THIS HAPPENED ON AN UNKNOWN DATE IN MAY 2022. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D9, H3, H6,H10 H10: THE DEVICE WAS RECEIVED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED, AND THE REPORTED ISSUE COULD NOT BE REPRODUCED. UPSTREAM OCCLUSION TESTING WAS PERFORMED AND AN UPSTREAM OCCLUSION WAS PROPERLY DETECTED. A SERVICE HISTORY REVIEW WAS PERFORMED AND REVEALED THAT THE DEVICE HAS NO PREVIOUS SERVICE EVENTS; THEREFORE, SERVICING DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THE REPORTED CONDITION WAS NOT VERIFIED. THERE IS CURRENTLY AN OPEN FIELD ACTION (FA-2021-056) ASSOCIATED WITH REINFORCING PROPER INTRAVENOUS LINE SETUP AND DETECTION OF UPSTREAM OCCLUSION FOR SPECTRUM PUMPS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT A SPECTRUM IQ PUMP WOULD NOT UPSTREAM OCCLUDE DURING PREVENTIVE MAINTENANCE CYCLE. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1472573 SPECTRUM IQ INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION NA NA 00085412610900

Patients

Seq Age Sex Outcome Treatment
1 Unknown