FDA Adverse Event Injury Summary report: N

ZOOM¿ 88 LARGE DISTAL PLATFORM

MDR report key: 14877123 · Received June 30, 2022

Report

Report Number
3014590708-2022-00012
Event Type
Injury
Date Received
June 30, 2022
Date of Event
May 27, 2022
Report Date
June 30, 2022
Manufacturer
IMPERATIVE CARE. INC
Product Code
QJP
UDI-DI
00812212030337
PMA / PMN Number
K212224
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

BASED ON THE INFORMATION PROVIDED, THE RETRACTION WAS REPORTED AS HARDER THAN NORMAL BASED ON HAVING TO PULL THROUGH THE PATIENT'S TORTUOUS ANATOMY. THE DEVICE WAS DISCARDED AFTER USE AND THEREFORE NOT AVAILABLE FOR RETURN AND INVESTIGATION. THERE WERE NO DEVICE DEFICIENCIES REPORTED RELATED TO THE TRACSTAR. THE LOT NUMBER WAS NOT PROVIDED BY THE USER FACILITY. ALL DEVICES UNDERGO TESTING AND INSPECTION CONFIRMING DEVICES MEET SPECIFICATIONS PRIOR TO RELEASE.

Description of Event or Problem · 0

THE PATIENT WAS UNDERGOING A VENOUS STENTING CASE. UPON COMPLETION OF THE CASE AND DURING RETRACTION, THE DISTAL PORTION OF THE TRACSTAR GOT CAUGHT ON THE PREVIOUSLY DEPLOYED STENT AND BROKE SEVERAL CENTIMETERS FROM THE DISTAL END OF THE CATHETER IN THE INTERNAL JUGULAR. THE PATIENT REPORTEDLY HAD A VERY ZIG-ZAG TORTUOUS ANATOMY. THE DISTAL PORTION OF THE CATHETER WAS RETRIEVED WITH A SNARE. PATIENT WAS REPORTED TO BE DOING FINE AND NO COMPLICATIONS WERE REPORTED. THE PHYSICIAN DID NOT REPORT ANY ISSUES WITH THE TRACSTAR. DURING FOLLOW-UP, THE PHYSICIAN INFORMED THAT THE SHORT SHEATH WAS IN THE "Z" OR ZIG-ZAG SHAPE IN THE BODY AS A FACTOR FOR THE TRACSTAR SHAFT BEING BROKEN. THE RETRACTION WAS HARDER THAN NORMAL BASED ON HAVING TO PULL THROUGH THIS TORTUOUS "Z".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2057971 ZOOM¿ 88 LARGE DISTAL PLATFORM GUIDE CATHETER QJP IMPERATIVE CARE. INC ICAC088095 00812212030337

Patients

Seq Age Sex Outcome Treatment
1 3 YR Female Required Intervention CAROTID STENT| GUIDEWIRE| SHORT SHEATH| SNARE