FDA Adverse Event Malfunction Summary report: N

COAGUCHEK ® XS SYSTEM

MDR report key: 14876894 · Received June 30, 2022

Report

Report Number
1823260-2022-01928
Event Type
Malfunction
Date Received
June 30, 2022
Date of Event
June 8, 2022
Report Date
June 30, 2022
Manufacturer
ROCHE DIAGNOSTICS
Product Code
GJS
UDI-DI
00365702127104
PMA / PMN Number
K062925
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER WAS ASSISTED WITH CHANGING THE UNIT OF MEASURE ON THE METER FROM %Q TO INR. THE REPORTER'S STRIPS WERE PROVIDED FOR INVESTIGATION WHERE THEY WERE TESTED USING A RETENTION METER AND CONTROLS. TESTING RESULTS (QC RANGE = 2.1 - 3.3 INR): QC 1: 3.0 INR, QC 2: 3.0 INR, QC 3: 3.0 INR. THE OBTAINED QC VALUES WERE IN THE ALLOWED RANGE OF THE USED COMBINATION STRIP LOT - QC LOT. ALL MEASUREMENTS WERE WITHOUT ERROR MESSAGES. ON A REGULAR BASIS, COAGUCHEK STRIPS OF LOTS CURRENTLY VALID IN THE MARKET ARE TESTED AS PART OF ROUTINE RETENTION TESTING AND RESULTS HAVE PASSED THE INTERNAL INSPECTION. PER PRODUCT LABELING, "COAGUCHEK USES HUMAN RECOMBINANT THROMBOPLASTIN. THEREFORE, THE COMPARABILITY TO OTHER HUMAN RECOMBINANT THROMBOPLASTINS IS BEST, WHEREAS HIGHER DEVIATIONS CAN OCCUR WITH OTHER THROMBOPLASTINS. HOWEVER, THOSE HIGHER DIFFERENCES BETWEEN THROMBOPLASTINS OF DIFFERENT (RABBIT, BOVINE BASED) ORIGIN ARE NOT A COAGUCHEK SPECIFIC ISSUE. SIMILAR DIFFERENCES CAN BE OBSERVED WHEN A HUMAN RECOMBINANT THROMBOPLASTIN-BASED LABORATORY METHOD IS COMPARED AGAINST SEVERAL OTHER (RABBIT, BOVINE-BASED) LABORATORY METHODS." OCCUPATION WAS PATIENT/CONSUMER.

Description of Event or Problem · 0

THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM COAGUCHEK XS METER SERIAL NUMBER (B)(4). AT 10:30 AM, THE RESULT FROM THE LABORATORY USING AN UNKNOWN REAGENT WAS 2.6 INR. AT 1:42 PM, THE RESULT FROM THE METER WAS 15%Q (3.4 INR). THE CUSTOMER IS ON A LOW SODIUM DIET. THE THERAPEUTIC RANGE WAS 2.5-3.5 INR AND THE CUSTOMER TESTS EVERY TWO WEEKS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2397462 COAGUCHEK ® XS SYSTEM PROTHROMBIN TIME TEST STRIPS GJS ROCHE DIAGNOSTICS NA 57790221 00365702127104

Patients

Seq Age Sex Outcome Treatment
1 81 YR Female ASPIRIN,| CALCIUM,| CARDIA,| IRON,| OMEPRAZOLE,| PROBIOTIC,| SERTRALINE,| SPIRONOLACTONE,| SYNTHROID,| TORSEMIDE.| TRELEG,| WARFARIN,