COAGUCHEK ® XS SYSTEM
Report
- Report Number
- 1823260-2022-01928
- Event Type
- Malfunction
- Date Received
- June 30, 2022
- Date of Event
- June 8, 2022
- Report Date
- June 30, 2022
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- GJS
- UDI-DI
- 00365702127104
- PMA / PMN Number
- K062925
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER WAS ASSISTED WITH CHANGING THE UNIT OF MEASURE ON THE METER FROM %Q TO INR. THE REPORTER'S STRIPS WERE PROVIDED FOR INVESTIGATION WHERE THEY WERE TESTED USING A RETENTION METER AND CONTROLS. TESTING RESULTS (QC RANGE = 2.1 - 3.3 INR): QC 1: 3.0 INR, QC 2: 3.0 INR, QC 3: 3.0 INR. THE OBTAINED QC VALUES WERE IN THE ALLOWED RANGE OF THE USED COMBINATION STRIP LOT - QC LOT. ALL MEASUREMENTS WERE WITHOUT ERROR MESSAGES. ON A REGULAR BASIS, COAGUCHEK STRIPS OF LOTS CURRENTLY VALID IN THE MARKET ARE TESTED AS PART OF ROUTINE RETENTION TESTING AND RESULTS HAVE PASSED THE INTERNAL INSPECTION. PER PRODUCT LABELING, "COAGUCHEK USES HUMAN RECOMBINANT THROMBOPLASTIN. THEREFORE, THE COMPARABILITY TO OTHER HUMAN RECOMBINANT THROMBOPLASTINS IS BEST, WHEREAS HIGHER DEVIATIONS CAN OCCUR WITH OTHER THROMBOPLASTINS. HOWEVER, THOSE HIGHER DIFFERENCES BETWEEN THROMBOPLASTINS OF DIFFERENT (RABBIT, BOVINE BASED) ORIGIN ARE NOT A COAGUCHEK SPECIFIC ISSUE. SIMILAR DIFFERENCES CAN BE OBSERVED WHEN A HUMAN RECOMBINANT THROMBOPLASTIN-BASED LABORATORY METHOD IS COMPARED AGAINST SEVERAL OTHER (RABBIT, BOVINE-BASED) LABORATORY METHODS." OCCUPATION WAS PATIENT/CONSUMER.
THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM COAGUCHEK XS METER SERIAL NUMBER (B)(4). AT 10:30 AM, THE RESULT FROM THE LABORATORY USING AN UNKNOWN REAGENT WAS 2.6 INR. AT 1:42 PM, THE RESULT FROM THE METER WAS 15%Q (3.4 INR). THE CUSTOMER IS ON A LOW SODIUM DIET. THE THERAPEUTIC RANGE WAS 2.5-3.5 INR AND THE CUSTOMER TESTS EVERY TWO WEEKS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2397462 | COAGUCHEK ® XS SYSTEM | PROTHROMBIN TIME TEST STRIPS | GJS | ROCHE DIAGNOSTICS | NA | 57790221 | 00365702127104 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Female | ASPIRIN,| CALCIUM,| CARDIA,| IRON,| OMEPRAZOLE,| PROBIOTIC,| SERTRALINE,| SPIRONOLACTONE,| SYNTHROID,| TORSEMIDE.| TRELEG,| WARFARIN, |