FDA Adverse Event Other Summary report: N

DA VINCI SURGICAL SYSTEM

MDR report key: 1487679 · Received September 18, 2009

Report

Report Number
2955842-2009-00312
Event Type
Other
Date Received
September 18, 2009
Date of Event
August 27, 2009
Report Date
August 27, 2009
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
PMA / PMN Number
K070684
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION CONDUCTED BY ISI FIELD SERVICE ENGINEERING CONCLUDED THAT THE ISSUE EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH A PATIENT MANIPULATOR (PSM). THE PSM IS AN INSTRUMENT ARM LOCATED ON THE PATIENT SIDE CART AND PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED PSM. THE PSM WAS RETURNED TO ISI FOR EVALUATION. ENGINEERING DISCOVERED THAT A PIECE OF THE CANNULA (LOCK/BEARING MECHANISM) HAD BROKEN OFF AND WAS RATTLING AROUND THE ELBOW AREA OF THE PSM. AS OF SEPTEMBER 17, 2009, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SURGICAL PROCEDURE, THE CUSTOMER DETECTED A MECHANICAL PROBLEM WITH ONE OF THE SYSTEM ARMS. THE SURGEON DECIDED TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO COMPLETE THE PLANNED PROCEDURE. NO PATIENT HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL, INC. IS1200 A4.3P9

Patients

Seq Age Sex Outcome Treatment
1 Other ACCESSORIES| DA VINCI SURGICAL SYSTEM INSTRUMENTS