DA VINCI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2009-00312
- Event Type
- Other
- Date Received
- September 18, 2009
- Date of Event
- August 27, 2009
- Report Date
- August 27, 2009
- Manufacturer
- INTUITIVE SURGICAL, INC.
- Product Code
- NAY
- PMA / PMN Number
- K070684
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE INVESTIGATION CONDUCTED BY ISI FIELD SERVICE ENGINEERING CONCLUDED THAT THE ISSUE EXPERIENCED BY THE CUSTOMER WAS ASSOCIATED WITH A PATIENT MANIPULATOR (PSM). THE PSM IS AN INSTRUMENT ARM LOCATED ON THE PATIENT SIDE CART AND PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED PSM. THE PSM WAS RETURNED TO ISI FOR EVALUATION. ENGINEERING DISCOVERED THAT A PIECE OF THE CANNULA (LOCK/BEARING MECHANISM) HAD BROKEN OFF AND WAS RATTLING AROUND THE ELBOW AREA OF THE PSM. AS OF SEPTEMBER 17, 2009, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING A DA VINCI SURGICAL PROCEDURE, THE CUSTOMER DETECTED A MECHANICAL PROBLEM WITH ONE OF THE SYSTEM ARMS. THE SURGEON DECIDED TO CONVERT TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO COMPLETE THE PLANNED PROCEDURE. NO PATIENT HARM WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL, INC. | IS1200 A4.3P9 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | ACCESSORIES| DA VINCI SURGICAL SYSTEM INSTRUMENTS |