FDA Adverse Event Injury Summary report: N

TITAN TOUCH

MDR report key: 14875660 · Received June 30, 2022

Report

Report Number
2125050-2022-00677
Event Type
Injury
Date Received
June 30, 2022
Date of Event
June 3, 2022
Report Date
June 30, 2022
Manufacturer
COLOPLAST A/S
Product Code
FHW
UDI-DI
05708932539203
PMA / PMN Number
P000006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. SHOULD ADDITIONAL FACTS PROMPT US TO ALTER OR SUPPLEMENT ANY INFORMATION OR CONCLUSIONS CONTAINED IN THE ORIGINAL MDR OR IN ANY PRIOR SUPPLEMENTAL REPORTS, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE LOT NUMBER (8151739) WAS REVIEWED FOR COMPLAINT TREND, NONCONFORMING REPORT AND CAPA. DEVICES MET SPECIFICATION PRIOR TO RELEASE AND NO TRENDS WERE NOTED.

Description of Event or Problem · 0

ACCORDING TO AVAILABLE INFORMATION, THIS DEVICE MALFUNCTIONED DUE TO SWELLING. 10CC OF FLUID WAS REMOVED FROM THE (L) HYDROCELE. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1029742 TITAN TOUCH INFLATABLE PENILE PROSTHESIS FHW COLOPLAST A/S ES29222400 8151739 05708932539203

Patients

Seq Age Sex Outcome Treatment
1 73 YR Male Required Intervention