FDA Adverse Event Malfunction Summary report: N

ACT PLUS INSTRUMENT

MDR report key: 14872992 · Received June 30, 2022

Report

Report Number
2184009-2022-00157
Event Type
Malfunction
Date Received
June 30, 2022
Date of Event
June 6, 2022
Report Date
October 11, 2022
Manufacturer
PERFUSION SYSTEMS
Product Code
GKN
PMA / PMN Number
K940426
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

UPDATE TO B5: DURING SERVICING OF THIS ACT PLUS INSTRUMENT, THE CUSTOMER'S BIO-MED TECHNICIAN OBSERVED THAT THE INSTRUMENT WAS HAVING A MEASUREMENT MALFUNCTION AND ONLY MEASURED IN ONE CELL. THE CUSTOMER STATED THAT SOMETIMES THERE WAS BIG DEFERENCE BETWEEN THE READINGS FOR THE CHANNELS. THIS WAS DETECTED DURING SERVICE SO THERE WAS NO ADVERSE PATIENT EFFECT. DEVICE EVALUATION SUMMARY: THE REPORTED THE ISSUE OF INSTRUMENT WAS HAVING A MEASUREMENT MALFUNCTION, IT ONLY MEASURED IN ONE CELL AND SOMETIMES THERE WAS BIG DEFERENCE BETWEEN THE READINGS FOR THE CHANNELS WAS NOT VERIFIED DURING DEPOT SERVICE. THE DEPOT SERVICE TECHNICIAN PERFORMED THE ACTTRAC CARTRIDGE TEST AND ELECTRICAL SAFETY TEST AND THE RESULTS WERE AS EXPECTED. THE DEPOT SERVICE TECHNICIAN COULD NOT CONFIRM THE ISSUE. THE DEPOT SERVICE TECHNICIAN CLEANED THE REAGENT DRIVE SLIDER BLOCK AND THE INSTRUMENT AS A PRECAUTION. PREVENTIVE MAINTENANCE WAS COMPLETED PER SPECIFICATIONS. CONCLUSION: AFTER INVESTIGATION, THE COMPLAINT IS NOT CONFIRMED FOR THE ACT PLUS INSTRUMENTS MEASUREMENT MALFUNCTION AND ONLY MEASURING IN ONE CELL. THE ISSUE WAS NOT VERIFIED DURING DEPOT SERVICE. NO PATIENT/CLINICAL SAFETY ISSUES WERE REPORTED. TRENDS FOR ISSUES WITH THIS INSTRUMENT ARE REVIEWED AT QUARTERLY QUALITY MEETINGS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION THAT AT AN UNSPECIFIED TIME THIS ACT PLUS INSTRUMENT HAD A MEASUREMENT MALFUNCTION. USE OF INSTRUMENT WAS UNSPECIFIED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1058145 ACT PLUS INSTRUMENT TIMER, CLOT, AUTOMATED GKN PERFUSION SYSTEMS ACT200

Patients

Seq Age Sex Outcome Treatment
1 Unknown