FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 14859012 · Received June 29, 2022

Report

Report Number
3004753838-2022-113468
Event Type
Malfunction
Date Received
June 29, 2022
Date of Event
June 20, 2022
Report Date
September 6, 2022
Manufacturer
DEXCOM, INC.
Product Code
QBJ
PMA / PMN Number
DEN170088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). INITIAL REPORTER STATE - QC.

Additional Manufacturer Narrative · 0

(B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE TEST WAS PERFORMED AND PASSED. NORDIC BLUETOOTH PAIRING TEST WAS PERFORMED AND FAILED. NO DATA WAS PROVIDED FOR EVALUATION. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS A DEFECTIVE TRANSMITTER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1135378 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. 9445-24 5298422

Patients

Seq Age Sex Outcome Treatment
1 54 YR Male