FDA Adverse Event Injury Summary report: N

ZILVER VENA VENOUS SELF-EXPANDING STENT

MDR report key: 14854479 · Received June 29, 2022

Report

Report Number
3005580113-2022-00070
Event Type
Injury
Date Received
June 29, 2022
Date of Event
June 2, 2022
Report Date
June 29, 2022
Manufacturer
COOK IRELAND LTD
Product Code
QAN
UDI-DI
10827002574493
Adverse Event
Yes
Product Problem
Yes
Report Source
Distributor report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

COMMON NAME - QAN. PRODUCT CODE - QAN.

Description of Event or Problem · 0

ACCORDING TO THE INITIAL REPORTER: A FEMALE PATIENT IN HER 60S WITH COMPRESSION OF THE COMMON ILIAC AND EXTERNAL ILIAC VEINS ON RIGHT SIDE UNDERWENT AN IVUS AND VENOGRAPHY PROCEDURE IN WHICH THE ZILVER VENA VENOUS SELF-EXPANDING STENT, G57449 AND G57448, WERE USED. ACCESS SITE WAS RIGHT GROIN AND A IVUS AND VENOGRAPHY WAS PERFORMED. THE PHYSICIAN SELECTED 16X140 VENA TO STENT FROM THE CONFLUENCE INTO THE EXTERNAL ILIAC VEIN ON RIGHT SIDE. NO PREDILATION WAS DONE. AFTER DEPLOYING STENT OPTIMALLY, PHYSICIAN ADVANCED A (ANOTHER MANUFACTURER'S) 16X60 BALLOON INTO THE STENT. THE TECH WHO WAS SCRUBBED IN STATES THE PHYSICIAN WAS NOT STEPPING ON FLUORO WHILE ADVANCING BALLOON, AND ONCE HE DID STEP ON FLUORO BOTH THE PHYSICIAN AND TECH SAW THAT THE STENT HAD ALMOST COMPLETELY, ABOUT 95%, MIGRATED INTO THE PATIENT¿S IVC (SUBJECT OF THIS REPORT). THEY BOTH AGREED THAT THE BALLOON CATHETER BUMPED THE CAUDAL EDGE OF THE STENT AND CAUSED IT TO MIGRATE SIGNIFICANTLY. PHYSICIAN ATTEMPTED TO USE THE BARD ATLAS BALLOON TO PULL THE STENT BACK DOWN INTO THE COMMON ILIAC, BUT WAS UNSUCCESSFUL. THEN, PHYSICIAN DEPLOYED ANOTHER 16X140 VENA INTO COMMON ILIAC, OVERLAPPING THE 2 STENTS A FEW CENTIMETERS. DUE TO STENT MIGRATION, A THIRD STENT, THE 16X100, WAS NEEDED TO EXTEND INTO THE EXTERNAL ILIAC. PHYSICIAN USED SAME 16X60 ATLAS BALLOON TO POST DILATE THE STENTS, AND AS HE WAS ADVANCING THE BALLOON INTO THE 16X100 STENT, THE BALLOON BUMPED INTO THIS STENT ALSO AND CAUSED IT TO JUMP ABOUT 2 CM BUT WAS ABLE TO USE THE BALLOON TO PULL THE STENT BACK DOWN TO ORIGINAL LANDING ZONE (REF: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2379234 ZILVER VENA VENOUS SELF-EXPANDING STENT QAN COOK IRELAND LTD G57449 C1910129 10827002574493

Patients

Seq Age Sex Outcome Treatment
1 Female Required Intervention