FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ SYRINGE

MDR report key: 14853933 · Received June 29, 2022

Report

Report Number
3003152976-2022-00298
Event Type
Malfunction
Date Received
June 29, 2022
Date of Event
May 24, 2022
Report Date
July 15, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INITIAL REPORTER TELEPHONE NUMBER (B)(6). A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 2110020, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. TEN RETAINED SAMPLES OF THE REPORTED LOT WERE USED FOR ADDITIONAL EVALUATION. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, THE STOPPERS WERE VERIFIED TO BE PROPERLY ASSEMBLED ON TO THE PLUNGER ROD, AND NO LEAK WAS IDENTIFIED. BASED ON OUR INVESTIGATION, WE CANNOT IDENTIFY A DEFINITIVE ROOT CAUSE AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD PLASTIPAK¿ SYRINGE 2 DIFFERENT SYRINGES SOLUTION LEAKED FROM BARREL PAST THE PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SOLUTION LEAKING THROUGH FROM BARREL PAST THE PLUNGER ON 2 DIFFERENT SYRINGES" THE CUSTOMER REPLIED THERE WAS NO VISIBLE DAMAGE TO THE BARREL. THE SAMPLE IS ALSO NOT AVAILABLE TO BE SENT IN.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD PLASTIPAK¿ SYRINGE 2 DIFFERENT SYRINGES SOLUTION LEAKED FROM BARREL PAST THE PLUNGER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "SOLUTION LEAKING THROUGH FROM BARREL PAST THE PLUNGER ON 2 DIFFERENT SYRINGES." THE CUSTOMER REPLIED THERE WAS NO VISIBLE DAMAGE TO THE BARREL. THE SAMPLE IS ALSO NOT AVAILABLE TO BE SENT IN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2366146 BD PLASTIPAK¿ SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2110020

Patients

Seq Age Sex Outcome Treatment
1 Unknown