BD PLASTIPAK¿ SYRINGE
Report
- Report Number
- 3003152976-2022-00299
- Event Type
- Malfunction
- Date Received
- June 29, 2022
- Date of Event
- May 25, 2022
- Report Date
- July 15, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 2110020, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. TEN RETAINED SAMPLES OF THE REPORTED LOT WERE USED FOR ADDITIONAL EVALUATION. THE PRODUCT WAS VISUALLY INSPECTED, NO DEFECTS OR DAMAGE WAS NOTED, THE STOPPERS WERE VERIFIED TO BE PROPERLY ASSEMBLED ON TO THE PLUNGER ROD, AND NO LEAK WAS IDENTIFIED. BASED ON OUR INVESTIGATION, WE CANNOT IDENTIFY A DEFINITIVE ROOT CAUSE AT THIS TIME.
IT WAS REPORTED THAT THE BD PLASTIPAK¿ SYRINGE EXPERIENCED LEAKAGE PAST THE STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SOLUTION LEAKING THROUGH FROM BARREL PAST THE PLUNGER. THERE WAS NO VISIBLE DAMAGE TO THE BARREL.
IT WAS REPORTED THAT THE BD PLASTIPAK¿ SYRINGE EXPERIENCED LEAKAGE PAST THE STOPPER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SOLUTION LEAKING THROUGH FROM BARREL PAST THE PLUNGER. THERE WAS NO VISIBLE DAMAGE TO THE BARREL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2017347 | BD PLASTIPAK¿ SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2110020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |