WESTMED LLC
Report
- Report Number
- 2028807-2022-00016
- Event Type
- Malfunction
- Date Received
- June 28, 2022
- Date of Event
- June 6, 2022
- Report Date
- June 28, 2022
- Manufacturer
- WESTMED LLC
- Product Code
- OFP
- UDI-DI
- 00709078008290
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
THE SPLITTING OF THE TUBE CAUSED THE PATIENT TO RE-INTUBATED. BASE ON THE REPORTED INFORMATION THE CRITERIA FOR REPORTING AN ADVERSE EVENT HAS BEEN MET
THE SPLITTING OF THE TUBE CAUSED THE PATIENT TO RE-INTUBATED. BASE ON THE REPORTED INFORMATION THE CRITERIA FOR REPORTING AN ADVERSE EVENT HAS BEEN MET. COMPLAINT CONFIRMED WITH THE PHOTOS PROVIDED IN THE RELATED COMPLAINTS AND RETURNED GOODS. THE ROOT CAUSE WAS DETERMINED TO BE THAT THE HOSE THAT FEEDS ONE OF THE PLASTIC RESINS INTO THE BLENDER WAS NOT PROPERLY CONNECTED. THIS RESULTED IN AN INCORRECT RATIO OF RESINS, LEADING TO THE TUBING BEING MORE RIGID AND MORE LIKELY TO CRACK. LAB RESULTS CONFIRMED THAT TUBING MANUFACTURED AFTER THE HOSE WAS FIXED WAS THE CORRECT RATIO OF PLASTIC RESINS. WE COMPLETED A HEALTH HAZARD EVALUATION AS PART OF THE RISK ASSESSMENT. THE RISK ASSESSMENT PERFORMED WITH RMA-20030 INDICATED A SEVERITY OF 9. WE INITIATED CAPA-00495 TO ADDRESS THIS ISSUE. WE REVIEWED THE COMPLAINT HISTORY IN OUR QUALITY MANAGEMENT SYSTEM RELATED TO THE AFFECTED PRODUCT FOR THE 24 MONTHS PRECEDING THE EARLIEST COMPLAINT RECEIVED. WE DID NOT RECEIVE ANY COMPLAINT FOR THIS PRODUCT PRECEDING COMPLAINT 07496. ALSO, WE ANALYZED THE FREQUENCY OF COMPLAINTS WE HAVE RECEIVED FOR THE CIRCUIT PRODUCT FAMILY. THERE IS NO TRENDING ISSUE FOR THE CIRCUIT PRODUCT FAMILY. WE EMAILED RESOLUTION TO THE CUSTOMER.
SPLITS/CRACKS IN TUBING THE CRNS EXTENDED THE TUBING AFTER INDUCTION AND AN OBVIOUS BREAK IN THE TUBING OCCURRED IN TWO PLACES. THIS WAS A SERIOUS SITUATION BECAUSE OF THE PATIENT'S COMORBIDITIES AND THE CONCERN FOR ETT IN ESOPHAGUS.
SPLITS/CRACKS IN TUBING THE CRNS EXTENDED THE TUBING AFTER INDUCTION AND AN OBVIOUS BREAK IN THE TUBING OCCURRED IN TWO PLACES. THIS WAS A SERIOUS SITUATION BECAUSE OF THE PATIENT'S COMORBIDITIES AND THE CONCERN FOR ETT IN ESOPHAGUS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2365647 | WESTMED LLC | 96" ADULT EXP,GSE,PAR WYE,(2)BV,LG.MASK,3LF BAG,3LF BAG,10'X.050" MM GSL, | OFP | WESTMED LLC | 9677 | UNKNOWN | 00709078008290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |