FDA Adverse Event Malfunction Summary report: N

INPEN MMT-105NNPKNA NOVO NORDISK PINK

MDR report key: 14838508 · Received June 27, 2022

Report

Report Number
3012822846-2022-00896
Event Type
Malfunction
Date Received
June 27, 2022
Date of Event
June 3, 2022
Report Date
July 3, 2022
Manufacturer
COMPANION MEDICAL INC
Product Code
FMF
UDI-DI
0000108620880003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

PER VISUAL INSPECTION: CAP NO LONGER STAYS ATTACHED. NO PHYSICAL DAMAGE TO CARTRIDGE HOLDER OR INPEN WAS NOTED. UNIT PAIRED SUCCESSFULLY TO COMMERCIAL APP. MY INPEN MENU DISPLAYED: 10 TEST VALUES OF 4.0U DIALED AND DOSED. ALL VALUES DISPLAYED ACCURATELY IN THE LOGBOOK. ALL DOSES WERE WRITTEN TO THE APP LOGBOOK WITHIN 1 MINUTE. INPEN RECEIVED WITH LEADSCREW 1/4 OF THE TRAVEL. RE-WOUND SCREW. NO DRAG WAS OBSERVED, THE SCREW WAS NOT BENT, ADVANCED WHEN DOSAGE KNOB WAS PRESSED DIALING A DOSAGE AND RETRACTED APPROPRIATELY. NO RESISTANCE WAS OBSERVED WHEN DOSING WITHOUT A CARTRIDGE INSTALLED. THE SCREW ADVANCED EVERY TIME 30.0U WAS DIALED AND DOSED UNTIL THE SCREW REACHED MAX EXTENSION. IN CONCLUSION: INPEN RECEIVED WORKING AS DESIGN. NO MALFUNCTIONS NOTED DURING TESTING THAT COULD AFFECT INSULIN DELIVERY. THE CUSTOMER COMPLAINT OF INPEN NOT WORKING COULD NOT BE CONFIRMED. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INPEN WAS BROKEN OR DAMAGED AND WAS NOT WORKING. CUSTOMER TRIED DIFFERENT NEEDLES THE CARTRIDGE OR INPEN WOULDN'T PRIME. TROUBLESHOOTING WAS PERFORMED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2018309 INPEN MMT-105NNPKNA NOVO NORDISK PINK SYRINGE, PISTON FMF COMPANION MEDICAL INC MMT-105NNPKNA B0048 0000108620880003

Patients

Seq Age Sex Outcome Treatment
1 57 YR Female