FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE PULMONARY SURFACE TOTAL CARE FRAME
MDR report key: 1483647
·
Received July 28, 2009
Report
- Report Number
- 1824206-2009-03921
- Event Type
- Malfunction
- Date Received
- July 28, 2009
- Date of Event
- July 9, 2008
- Report Date
- July 9, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
TECHNICIAN REPLACED (GCI) & LEFT (UCM) CABLE TO RESOLVE THE ISSUE. PURSUANT TO OUR RESPONSE TO WARNING LETTER 2008-DT-04, HILL-ROM IS CONTINUING ITS RETROSPECTIVE REVIEW OF EVENTS FOR REPORTABILITY. THIS EFFORT WILL CONTINUE UNTIL WE ARE CURRENT.
Description of Event or Problem · 1
(GCI) FUNCTIONS VERY ERRATIC, LEFT (UCM) CABLE CUT - WIRES EXPOSED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE PULMONARY SURFACE TOTAL CARE FRAME | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM RITTER | P1900H006263 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |