FDA Adverse Event Malfunction Summary report: N

INSPIRE 8 HOLLOW FIBER OXYGENATOR

MDR report key: 14828581 · Received June 27, 2022

Report

Report Number
9680841-2022-00028
Event Type
Malfunction
Date Received
June 27, 2022
Date of Event
May 26, 2022
Report Date
November 2, 2022
Manufacturer
SORIN GROUP ITALIA SRL
Product Code
DTZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PATIENT INFORMATION WERE NOT PROVIDED. THE EXPIRATION DATE REFERS TO THE STERILE FINISHED PRODUCT. THE COMPLAINED INSPIRE 8 START OXYGENATOR (CATALOG NUMBER 050710, LOT 2108240165) IS NOT DISTRIBUTED IN THE USA, THEREFORE THE UDI IS NOT APPLICABLE. THE OXYGENATOR ITEM 050710 IS SIMILAR TO THE INSPIRE 8 OXYGENATOR 050714, WHICH IS DISTRIBUTED IN THE USA, FOR WHICH THE DEVICE IDENTIFIER IS (B)(4). PMA/510K. THE PRODUCT ITEM 050710 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO THE INSPIRE 8 OXYGENATOR 050714, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K130433). THE DEVICE MANUFACTURE DATE REFERS TO MANUFACTURE DATE OF THE STERILE, FINISHED OXYGENATOR. DEVICE MANUFACTURE DATE. SORIN GROUP ITALIA MANUFACTURES THE INSPIRE 8 START HOLLOW FIBER OXYGENATOR WITH INTEGRATED HARDSHELL VENOUS/CARDIOTOMY RESERVOIR. THE INCIDENT OCCURRED IN (B)(6). THE INVOLVED OXYGENATOR IS AVAILABLE AND HAS NOT BEEN REQUESTED. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS OBTAINED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 0

DURING FOLLOW UP WITH THE CUSTOMER, LIVANOVA WAS INFORMED THAT THE CHANGEOUT TOOK LESS THAN ONE MINUTE, SO THAT IT COULD PASS BLOOD VOLUME, THE HEART WAS MASSAGED WHILE IT WAS BEING CARRIED OUT.

Additional Manufacturer Narrative · 0

MEDICAL TEAM ELECTED TO CHANGE-OUT THE DEVICE AND PROCEDURE WAS COMPLETED WITHOUT ANY PATIENT IMPACT. THE CHANGEOUT TAKES LESS THAN ONE MINUTE, SO THE EVENT HAS BEEN REASSESSED AS NOT REPORTABLE. HOWEVER, AN INITIAL REPORT TO FDA WAS ALREADY SUBMITTED, THEREFORE, A FOLLOW UP HAS BEEN TRIGGERED. THE OXYGENATOR WAS RETURNED TO MANUFACTURER FOR PRODUCT ANALYSIS: THE REPORTED EVENT WAS NOT REPRODUCED AND DEVICE WAS FOUND TO BE PROPERLY WORKING AS PER DESIGN SPECIFICATION WITH NO DEVIATION IDENTIFIED. ANALYSIS OF PROVIDED PUMP-SHEET COULD NOT BE PERFORMED SINCE NO ALL THE RELEVANT DATA WERE MADE AVAILABLE. NO FURTHER SIMILAR COMPLAINTS HAVE BEEN RECORDED FOR NOTICED PRODUCT LOT ON A TOTAL OF 144 MANUFACTURED UNITS WORLDWIDE DISTRIBUTED, THUS EXCLUDING ANY SYSTEMATIC QUALITY PROBLEM. MANUFACTURING RECORDS HAVE BEEN VERIFIED TO BE CONFORMING TO SPECIFICATIONS. BASED ON HISTORICAL DATA ANALYSIS, THE HIGH TRANSMEMBRANE OXYGENATOR PRESSURE GRADIENT IS RELATED TO THE INCREASE OF HYDRAULIC RESISTANCE AS A RESULT OF UNDESIRED CELLULAR ACTIVATION ASSOCIATED WITH PLATELET ADHESION AND FIBRIN LAYER DEPOSITION INSIDE THE OXYGENATOR. THIS TYPE OF EVENT IS A MULTI-FACTORIAL PHENOMENON POSSIBLY TRIGGERED BY A COMBINATION OF CLINICAL PROCEDURE, THERAPIES (E.G. ANTICOAGULANT PRESCRIPTION, HEPARIN COMPOSITION AND PRIMING COMPOSITION) AND PATIENT-SPECIFIC HEALTH CONDITIONS. NO SPECIFIC ACTION WAS CURRENTLY DEEMED NECESSARY. LIVANOVA WILL KEEP MONITORING THE MARKET.

Description of Event or Problem · 0

SORIN GROUP ITALIA HAS RECEIVED A REPORT THAT, AT THE BEGINNING OF A CEC WITH TEMPERATURE OF 36°C, AFTER PURGING THE CARDIOPLEGIA, NO FLOW COULD BE ACHIEVED. MEDICAL TEAM ELECTED TO CHANGE-OUT THE OXYGENATOR. THERE IS NO REPORT OF ANY PATIENT INJURY.

Description of Event or Problem · 0

SEE INITIAL REPORT.

Description of Event or Problem · 0

SEE INTIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1055599 INSPIRE 8 HOLLOW FIBER OXYGENATOR OXYGENATOR, CARDIOPULMONARY BYPASS DTZ SORIN GROUP ITALIA SRL 2108240165

Patients

Seq Age Sex Outcome Treatment
1 64 YR Male