FDA Adverse Event Malfunction Summary report: N

LIBRE SENSOR FREESTYLE

MDR report key: 14826698 · Received June 27, 2022

Report

Report Number
2954323-2022-21103
Event Type
Malfunction
Date Received
June 27, 2022
Date of Event
June 20, 2022
Report Date
December 15, 2022
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
PZE
PMA / PMN Number
P160030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

REPEATED ATTEMPTS BY ADC TO RETRIEVE THE PRODUCT WERE UNSUCCESSFUL AND/OR THE CUSTOMER DISCARDED THE PRODUCT. AT THIS TIME PRODUCT HAS NOT YET BEEN RETURNED. AN EXTENDED INVESTIGATION HAS BEEN PERFORMED FOR THE REPORTED COMPLAINT AND THERE WAS NO INDICATION THAT THE PRODUCT DID NOT MEET SPECIFICATION. DHRS FOR THE FS LIBRE SENSOR AND FS LIBRE SENSOR KIT WERE REVIEWED AND THE DHRS SHOWED THE FS LIBRE SENSOR AND SENSOR KIT PASSED ALL TESTS PRIOR TO RELEASE. IF THE PRODUCT IS RETURNED, A PHYSICAL INVESTIGATION WILL BE PERFORMED AND A FOLLOW-UP REPORT SUBMITTED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

SENSOR (B)(6) HAS BEEN RETURNED AND INVESTIGATED. THE SENSOR PLUG WAS FULLY SEATED AND NO ISSUES WERE OBSERVED. EXTRACTED DATA FROM THE RETURNED SENSOR USING APPROVED SOFTWARE. THE SENSOR WAS FOUND TO BE IN SENSOR STATE 5 (INDICATING NORMAL TERMINATION). CORROSION WAS OBSERVED ON PLUG ASSEMBLY EXTENDED INVESTIGATION HAS BEEN PERFORMED. PQE PHYSICALLY INSPECTED THE RETURNED PUCK AND REVIEWED THE ATTACHMENT FROM THE PHASE 1 INVESTIGATION, AND OBSERVED CORROSION IN THE PLUG TRIANGLE AREA. PQE DETERMINED THE CORROSION OBSERVED ON THE PCBA TRIANGLE AREA IS NOT A CONTRIBUTING FACTOR TO THIS COMPLAINT. PQE DETERMINED THE CORROSIONS OBSERVED WERE NOT SEVERE ENOUGH TO HAVE ANY IMPACT ON THE SENSORS ACCURACY. CURRENT WAS APPLIED TO THE SENSOR TO PERFORM ACCURACY TESTING WHILE IN THE TEST FIXTURE. ALL RESULTS WERE WITHIN SPECIFICATION. POISE VOLTAGE AND SENSOR THERMISTOR TESTING WERE BOTH WITHIN SPECIFICATION, INDICATING THE SENSOR WAS PROVIDING ACCURATE GLUCOSE READINGS. THEREFORE, THE ISSUE IS NOT CONFIRMED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 0

A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
697916 LIBRE SENSOR FREESTYLE FLASH GLUCOSE MONITORING SYSTEM PZE ABBOTT DIABETES CARE INC 71536-01

Patients

Seq Age Sex Outcome Treatment
1 Unknown