FDA Adverse Event Malfunction Summary report: N

MTS INCUBATOR

MDR report key: 1481853 · Received July 17, 2009

Report

Report Number
1056600-2009-00150
Event Type
Malfunction
Date Received
July 17, 2009
Date of Event
June 25, 2009
Report Date
July 17, 2009
Manufacturer
MICRO TYPING SYSTEMS, INC.
Product Code
JTQ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER DISCARDED THE CARDS THAT WERE IN THE INCUBATOR WHEN ISSUE WAS OBSERVED. INVESTIGATION CONDUCTED REGARDING THE RETURNED MTS INCUBATOR. THE INCUBATOR TEMPERATURE IS WITHIN SPECIFIED RANGE. THE CUSTOMER'S CONCERN WAS NOT CONFIRMED. THE CUSTOMER RECEIVED A REPLACEMENT INCUBATOR TO RESOLVE THE ISSUE. (B) (4)

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE TEMPERATURE OF THE MTS INCUBATOR WAS OUT OF SPECIFICATION. NO ERRONEOUS RESULTS WERE REPORTED. A TEMPERATURE OUT OF SPECIFICATION, IF UNDETECTED DURING TESTING MAY LEAD TO ERRONEOUS TEST RESULTS AND MAY LEAD TO THE TRANSFUSION OF INCOMPATIBLE BLOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MTS INCUBATOR MTS GEL CARD INCUBATOR JTQ MICRO TYPING SYSTEMS, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1