FREESTYLE FREEDOM LITE
Report
- Report Number
- 2954323-2009-01740
- Event Type
- Injury
- Date Received
- September 18, 2009
- Date of Event
- August 18, 2009
- Report Date
- September 18, 2009
- Manufacturer
- ABBOTT DIABETES CARE INC., USA
- Product Code
- LFR
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NO INFORMATION
Narratives
THIS IS A FINAL REPORT. THE DEVICE AND TEST STRIPS (LOT NO: 0913101) WERE RETURNED AND AN INVESTIGATION USING APPROVED CONTROL SOLUTION DETERMINED THE COMPLAINT IS NOT CONFIRMED. ALL RESULTS WERE WITHIN RANGE SPECIFICATION DURING CONTROL SOLUTION TESTING. A REVIEW OF THE METER'S INTERNAL MEMORY LOG REVEALED A RESULT OF 90 MG/DL IN 2009. IT SHOULD BE NOTED: WHILE TROUBLESHOOTING WITH CUSTOMER SERVICE, IT WAS DISCOVERED THE CUSTOMER DID NOT WASH PRIOR TO TESTING WHICH MAY CAUSE IMPRECISE READINGS.
CUSTOMER REPORTED WAKING UP AT 2:30 AM IN 2009, AND USING HER WALKER, FELL, INJURING HER KNEE, WHILE WALKING TOWARDS THE BATHROOM. SHE NOTED SHE HAD NO OTHER SYMPTOMS TO REPORT. SHE DENIED FEELING LIGHTHEADED, WEAK OR NAUSEOUS. SHE FURTHER NOTED HER HUSBAND CHECKED HER GLUCOSE USING HER FREESTYLE FREEDOM LITE BLOOD GLUCOSE METER AND RECEIVED A RESULT OF 90 MG/DL. PARAMEDICS WERE CALLED AND THEY RECEIVED A RESULT OF 42 MG/DL, APPROXIMATELY 30 TO 45 MINUTES LATER. CUSTOMER INITIALLY REFUSED TO BE TRANSPORTED TO A LOCAL EMERGENCY ROOM BECAUSE SHE NOTED SHE FELT FINE. SHE LATER AGREED AND WAS GIVEN "SUGAR STUFF" BY THE PARAMEDICS. AT THE HOSPITAL SHE WAS DIAGNOSED WITH HYPOGLYCEMIA AND GIVEN ADDITIONAL GLUCOSE VIA INTRAVENOUS INFUSION. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM LITE | BLOOD GLUCOSE MONITORING SYSTEM | LFR | ABBOTT DIABETES CARE INC., USA | NI | 0913101 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | Hospitalization| R |