FDA Adverse Event Malfunction Summary report: N

AB MEDICA CURVED SCISSOR AND INSERT

MDR report key: 14801183 · Received June 23, 2022

Report

Report Number
MW5110491
Event Type
Malfunction
Date Received
June 23, 2022
Date of Event
June 16, 2022
Report Date
June 21, 2022
Manufacturer
AB MEDICA SAS
Product Code
GCJ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

THE REQUALIFICATION OF OUR THERMOSEALING PROCESS SHOWED A NON CONFORMITY ON OUR SEALINGS. AFTER FURTHER INVESTIGATION, THE TYVEK BATCH USED IN THESES SEALINGS HAS BEEN FOUND NON CONFORMANT BECAUSE THE SEAL WIDTH IS SMALLER THAN EXPECTED. THE EVALUATION OF THE NON-CONFORMITY SHOWED THAT THIS BATCH HAS BEEN USED IN PRODUCT PACKAGING DISTRIBUTED IN THE US. THIS DEFECT IS NOT AFFECTING THE STERILE STATE OF THE PRODUCT SINCE THE INTEGRITY TESTING ACCORDING TO ASTM F 1929 HAS BEEN FOUND CONFORM. FOR THE FIRST TEST, 9 PRODUCTS WERE FOUND NON CONFORMANT BY THE LABORATORY RESCOLL FINDING THAT THE SEAL WIDTH WAS 3 TO 5.5MM ON THOSE 9 PRODUCTS WHEREAS IT IS REQUIRED AT LEAST 6 MM. THE SECOND TEST REALIZED IN THE SAME LABORATORY WAS CONFORM ON ALL 27 PRODUCTS TESTED DEMONSTRATING THAT THE PACKAGING REMAINS SEALED. FDA SAFETY REPORT ID# (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1267215 AB MEDICA CURVED SCISSOR AND INSERT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ AB MEDICA SAS LIL-IP-2SC-20ST
1267216 AB MEDICA CURVED SCISSOR AND INSERT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ AB MEDICA SAS LIL-IP-2SC-20ST
1267217 AB MEDICA CURVED SCISSOR AND INSERT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ AB MEDICA SAS LIL-IP-2SC-20ST
1267218 AB MEDICA CURVED SCISSOR AND INSERT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ AB MEDICA SAS LIL-IP-2SC-20ST
1267219 AB MEDICA CURVED SCISSOR AND INSERT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ AB MEDICA SAS LIL-IP-2SC-20ST
608918 AB MEDICA CURVED SCISSOR AND INSERT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ AB MEDICA SAS LIL-IP-2SC-20ST
608919 AB MEDICA CURVED SCISSOR AND INSERT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ AB MEDICA SAS LIL-IP-2SC-20ST
608920 AB MEDICA CURVED SCISSOR AND INSERT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ AB MEDICA SAS LIL-IP-2SC-20ST
608921 AB MEDICA CURVED SCISSOR AND INSERT LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ AB MEDICA SAS LIL-IP-2SC-20ST EK07210008

Patients

Seq Age Sex Outcome Treatment
1 Unknown